3
3. Hold the instrument vertically over the specimen.
Apply the instrument foot rapidly but without shock.
Apply just sufficient force to obtain firm contact between
instrument foot and the specimen. Hold for 1 sec and
max value shows on the display.
4. Next measurement. Push the ZERO-button and
repeat step 3 above.
Max measuring mode can be cancelled by press MAX-
button again. In such case the reading is an instant
value.
Average value
Average value can given from 2 – 9 measurements.
1. Push the power Ө-button.
2. Push the MAX-button. MAX shows on the display.
3. Push and release the N/Ave-button. Continue push
and release N/Ave-button to select numbers of
measurment in the calculation. Wait 3 secunds to the
display shows 0.
4. Perform first measurments as above.
5. Pushet ZERO-button after each measurement till all
measurements are made.
Display shows AVE symbolen, average value and
numbers of measurments.
Calibration
1. Hold the instrument vertical and make sure the probe
are free. The displayen show 0. If not push the ZERO-
button.
English