background image

Non Contact

Forehead

Thermometer

Specifications

Warnings

Temperature measurement range:

 

Forehead mode: 93.2~108°F (34~42.2°C)

 

Surface mode: -7.6~176°F (-22~80°C)

Operating temperature range: 

 

50~104°F (10~40°C), 15%~85% RH

Storage temperature range: 

 

It should be stored at room temperature 

 

between -4~122°F  (-20~50°C) , RH

85%

Transportation temperature 

 

less than 70°C, RH

95%

Atmospheric pressure: 

 800~1013 

hPa 

Accuracy:

 

Forehead mode: ±0.4°F (0.2°C) within 95~107.6°F 

 

(35~42°C), ±0.5°F (0.3°C) for other range.

 

Surface mode: ±0.5°F (0.3°C) within 71.6~108°F 

 

(22~42.2°C), others ±4% or ±4°F (2°C) whichever 

 is 

greater.

Fever alarm, memory and °F/°C switch function

Battery: one lithium cell battery (CR2032 *1pcs)

Battery life: around 3,000 continuous readings

Expected Service Life: 4 years

Dimensions: 80.1 x 46.4 x 37.7 mm

Weight: 46.3 grams including battery

• The device should not submerge into any liquids and 

  expose it to direct moisture.

• There is no gender and age limitation for using the 

 infrared 

thermometer.

• This is not an AP or APG product

USER MANUAL

Functions

Forehead Temperature

The thermometer has been designed for practical 

use. It’s not meant to replace a visit to the doctor. 

Please also remember to compare the 

measurement result to your regular body 

temperature. Please consult with doctor if you 

have health concerns.

Please see the “Use of the thermometer” section 

to learn how to measure the body temperature.

Surface Temperature

The surface mode shows the actual and 

unadjusted surface temperature which is different 

from the body temperature. It can help you 

monitor if the object temperature is suitable for the 

baby or patient, for example the baby’s milk.

Please see the “Use of the thermometer” section 

to learn how to measure the object temperature.

Room Temperature

Suitable ambient temperature is important for the 

baby and patient. The thermometer always helps 

you recognize the room temperature.

After power off, the room temperature will be 

shown on the screen with      icon.

Fever Alarm

If the thermometer detects a temperature 

99.5°F 

(or 37.5°C) under forehead mode, three short 

beep sound will follow one long beep sound to 

warn the user for potential fever.

Last Reading

When you get a new body temperature, the last 

reading will be shown on the screen (at top right 

corner) with the       icon.

Note: The device must stay in stable ambient (room) 

temperature for 15 minutes before operating.

1.   Always make sure the probe lens is clean without any 

 

damage and the forehead is clean.

Warning: Choking from swallowing small parts and batteries 

by children or pets is possible, please keep small parts and 

batteries at places where children and pets can’t reach.

2.   Power on:

 

For the first time, pull out the insulated strip from the 

 

battery cover to activate the thermometer.

 

Press the       ON/MEM button. 

Functions

Use

Memory Locations

There are total 25 sets of measurement condition 

for body temperature.

When power on, press the       ON/MEM button to 

see the temperature records with       icon.

°F / °C Switch

In “Power Off” mode, press and hold the “SCAN” button, then press 

the       ON/MEM button for 3 seconds, icon “°F” will be switched to 

icon “°C”. You can also use the same process to change the LCD 

display from °C to °F.