7
Sabina
™
II sit-to-stand lift • 7EN155106 Rev 1
www.hill-rom.com I www.liko.com
5. Connect the cables as follows:
- 1. cable, lift motor.
- 2. cable, motor for base-width
adjustment.
- H. cable, hand control.
9. Install the lower leg support on the lift mast. Loosen
wheel A in order to adjust the distance to the patient’s
lower legs. Loosen wheel B in order to adjust the
height setting. After adjustment, lock wheels A and B.
8. A) Put the frame for the foot rest over the lift mast’s bracket
to the base. Make sure the frame is securely in place.
B)
Push the footrest into the frame.
10. Release the emergency stop
by turning the button in the
direction indicated by the
arrows on the button.
11.
If applicable:
Attach the holder for quick
reference guide on the lift mast
according to the assembly
instruction. Place the quick
reference guide in the holder.
• the motion of the lift arm corresponds to the
buttons on
the hand control
• the emergency lowering device works (mechanically and
electrically)
After assembly, make sure that:
•
the base-width adjustment works
• the wheel brakes are working
• the indicator lamps on the front of the control box
illuminate during charging.
ON CHARGE
EMERGENCY
BATTERY
A
B
S
MS
M
L
XL
XXL
12. Before the first use, the lift’s
battery should be charged for
at least 6 hours. For detailed
instructions, see ”Charging the
Batteries”, page 9.
A
B