background image

Golvo

 

ServiceManual Golvo™ mobile lift (3EN400404 Rev 3)  

 

55

 

www.liko.com

  

Easy Strapper G7/G8, G77/G88

Monteringssats komplett lyftband G7/G8, G77/G88

Assembly Instruction

Monteringsanvisning

 

Prod. No. 20090069, 20090070 

3SE640248-03

2011-03-30

© Copyright Liko AB 

www.liko.com

1.

2.

3.

4.

 Mount the spare part, 

  assemble in reverse order.

 

 NOTE

  - After changing lift strap the Limit Switch must be tested by lifting maximum load all the way up to the highest position

  

through the entire lifting range. 

  - With the emergency stop out, move the liftstrap all the way up against the Limit switch. Check that the motor does not   

  operate when touching the Limit switch. 

  - Make sure the spare part is mounted according to the assembly instruction. 

  - Check all functions related to the changed spare part. 

  - Repairs and maintenance may only be performed by personnel authorized by Hill-Rom/Liko and using original Liko spare parts.

20090069 Contains:   Qty:  Fits: 

Easy strapper G7/G8 

Golvo 7000

 

 

Golvo 7000 ES Classic

 

 

Golvo 8000 (all models) 

  

20090070 Contains:   Qty:  Fits: 

Easy strapper G77/G88 

Golvo 7007 from S/N 7723630

 

 

Golvo 7007 ES Classic

 

 

Golvo 7007 LowBase

 

 

Golvo 8008 (all models) 

Now with grey strap and  

grey plastic details.

Tools required:

3mm

13mm

5.323-324

Summary of Contents for Golvo G7

Page 1: ...ut move the liftstrap all the way up against the Limit switch Check that the motor does not operate when touching the Limit switch Make sure the spare part is mounted according to the assembly instruc...

Page 2: ...m Contains Qty Coverage list Lowrider 2 Leg alu right 1 Leg alu left 1 Leg adapter 2 Wheel housing Golvo LowR 2 Wheel axle 12x64 Golvo Lo 2 Leg LowRider 2 Wheel D50 Golvo LowRider 4 Leg clip 2 Bearing...

Page 3: ...o AB www liko com 5 330 3 Assemble the rear wheels with stop screw 2 Assemble the legs with screws nuts plastic covers and washers from the lift NOTE Make sure the spare part is mounted according to t...

Page 4: ...y your hand control If other lifts are affected by your hand control Use the large UP and DOWN buttons on the operation panel of the lift that is to be used Press both buttons simultaneously until a s...

Page 5: ...006 with CSA C22 2 No 60601 1 08 Int Op 10 90 Active time 2 min 7 330925 000001 8301209 24 V DC 2 2 Ah IP54 IEC 60601 1 TESTED EAN 13 Serial no Y M 2012 3 Max xxx kg xxx lbs 9801154 Product name Prod...

Page 6: ...trol IR about 20 cm 0 8 inch from the control unit s IR receiver and press the large UP and DOWN buttons on the hand control simultaneously for about 2 seconds A signal is heard when the function is a...

Page 7: ...0925 000001 8301209 24 V DC 2 2 Ah IP54 IEC 60601 1 TESTED EAN 13 Serial no Y M 2012 3 Max xxx kg xxx lbs 9801154 Product name Prod No 1234567 www liko com Liko AB Nedre v gen 100 975 92 Lule Sweden S...

Page 8: ...ed according to the assembly instruction Check all functions related to the changed spare part Repairs and maintenance may only be performed by personnel authorized by Hill Rom Liko and using original...

Page 9: ...pare part is mounted according to the assembly instruction Check all functions related to the changed spare part Repairs and maintenance may only be performed by personnel authorized by Hill Rom Liko...

Page 10: ...mounted according to the assembly instruction Check all functions related to the changed spare part Repairs and maintenance may only be performed by personnel authorized by Hill Rom Liko and using ori...

Page 11: ...pyright Liko AB www liko com Tools required NOTE Make sure the spare part is mounted according to the assembly instruction Check all functions related to the changed spare part Repairs and maintenance...

Page 12: ...04 Copyright Liko AB www liko com Tools required 5 352 NOTE Make sure the spare part is mounted according to the assembly instruction Check all functions related to the changed spare part Repairs and...

Page 13: ...0010 Liko M230 2050015 Viking XL 2040013 and 2040033 Viking L 2040014 and 2040034 Viking M 2040015 and 2040035 NOTE Make sure the spare part is mounted according to the assembly instruction Check all...

Page 14: ...Qty HandControl hanger 10 132 mm 5 inch NOTE Make sure the spare part is mounted according to the assembly instruction Check all functions related to the changed spare part Repairs and maintenance may...

Page 15: ...required Contains Qty Lift strap joint 1 Strap joint plastic cover 1 Lift strap safety pin 4 mm 1 Fits Extension belts Golvo 7000 7007 Golvo 8000 8008 Likorall Multirall 200 www liko com Copyright Lik...

Page 16: ...e tested by lifting maximum load all the way up to the highest position Make sure the spare part is mounted according to the assembly instruction Check all functions related to the changed spare part...

Page 17: ...E640237 02 2010 05 10 Tools required Contains Qty Lift strap joint 1 Strap joint plastic cover 1 Lift strap safety pin 4mm 1 Fits Golvo 7000 7007 Golvo 8000 8008 Likorall www liko com Copyright Liko A...

Page 18: ...he assembly must be tested by lifting maximum load all the way up to the highest position Make sure the spare part is mounted according to the assembly instruction Check all functions related to the c...

Page 19: ...n plate 3 2 mm 20 Friction plate 3 4 mm 20 Tools required Loctite 406 Glue 1 When changing Spare part 20090014 Actuator replacement of Friction plate pos1 may be needed Use the same dimension as origi...

Page 20: ...changed spare part Repairs and maintenance may only be performed by personnel authorized by Hill Rom Liko and using original Liko spare parts When changing Spare part 20090038 Gear Rack set replace me...

Page 21: ...tikettstorlek 52 X 21 mm Layout With permanent marker fill in the S N at the product label same as on the Golvo 2 Conforms to ISO 10535 2006 with ANSI AAMI ES60601 1 2005 and is certified to ISO 10535...

Page 22: ...control unit s IR receiver and press the large UP and DOWN buttons on the hand control simultane ously for about 2 seconds A signal is heard when the function is activated The lift is now ready to use...

Page 23: ...AMI ES60601 1 2005 and is certified to ISO 10535 2006 with CSA C22 2 No 60601 1 08 Int Op 10 90 Active time 2 min 7 330925 000001 8301209 24 V DC 2 2 Ah IP54 IEC 60601 1 TESTED EAN 13 Serial no Y M 20...

Page 24: ...is mounted according to the assembly instruction Check all functions related to the changed spare part Repairs and maintenance may only be performed by personnel authorized by Hill Rom Liko and using...

Page 25: ...ls Uno with S N from 31825 Viking all models LikoLight Combine only with LikoScale all models Tools required N dv ndiga verktyg 6 mm 17 mm 1 x 1 x 10 mm 2 x 17 mm 2 x Golvo Uno Viking LikoLight WARNIN...

Page 26: ...k that the motor does not operate when touching the Limit switch Make sure the spare part is mounted according to the assembly instruction Check all functions related to the changed spare part Repairs...

Page 27: ...ls 4 mm 5 Remove Control unit Cable red blue Remove Actuator Cable black red 6 NOTE With the emergency stop out move the mast all the way up against the Limit switch Check that the motor does not oper...

Reviews: