Lighthouse REMOTE-4PN Series Operating Manual
3-14
248083317-1 Rev 4
Pump Remote Cable Install
1. Figure 3-20 shows the Pump Remote cable connector. The
connector block must be removed before attaching 22-16 AWG
stranded wires.
Figure 3-20 Pump Remote Cable Connector
Note:
If the remote
switch is no longer needed,
reinstall the jumper or the
pump will not run when
power is applied to the
REMOTE-4PN.
2. Loosen the two mounting screws using a small slotted screw driver.
Pull out to remove the connector block, then loosen the two screws
holding the wire jumper and replace it with 22-16 AWG twisted-
pair wire of sufficient length to reach the remote switch location.
3. Tighten the screws to hold the wire in place and reattach the
connector block by pushing in on the block. Make sure it is
properly positioned as it is keyed to fit only one way.
4. Tighten the two mounting screws to lock the block in place.
Energize Connections
1. Connect the signal cable to the hub or data-gathering equipment.
2. Connect the AC Power cord to the AC source.
3. Apply AC power by turning the power switch ON.
4. Power should be applied to the instrument, indicated by the
REMOTE-4PN Power LED illuminating; the pump should start
running if the jumper is in place or the Pump Remote switch is ON.
After approximately twenty seconds, data will be supplied by the
REMOTE-4PN Series to data-gathering equipment or network.
Best practices dictate that power to equipment on both ends of the data cable
should be OFF before a cable is attached. That is not always practical,
especially if a Lighthouse LMS Server or other equipment is monitoring
devices around the clock and service cannot be interrupted.
Mounting
Screws
Block
Summary of Contents for Remote 3014PN
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE 4 PN Series Airborne Particle Counter Operating Manual...
Page 6: ......
Page 10: ...Lighthouse REMOTE 4PN Series Operating Manual t iv 248083317 1 Rev 4...
Page 12: ...Lighthouse REMOTE 4PN Series Operating Manual ii 248083317 1 Rev 4...
Page 58: ...Lighthouse REMOTE 4PN Series Operating Manual 4 20 248083317 1 Rev 4...
Page 66: ...Lighthouse REMOTE 4PN Series Operating Manual A 2 248083317 1 Rev 4...
Page 78: ...Lighthouse REMOTE 4PN Series Operating Manual B 12 248083317 1 Rev 4...
Page 82: ...Lighthouse REMOTE 4PN Series Operating Manual I 4 248083317 1 Rev 4...
Page 83: ...00...
Page 84: ......