Getting Started
248083413-1 Rev 4
3-5
DIP Switch
Settings
DIP Switch General Definitions
Table 3-1 provides the DIP switch settings for each position, including
their descriptions. Each switch can be set in the following positions:
OFF (UP) = 0, ON (DOWN) = 1
Note:
Use a tool with a
very small pointed tip to
change DIP Switch
positions.
DIP Switch Addressing
Note:
Because Address
0 is reserved for
broadcasting in MODBUS
RS-485 communications,
Address 1 is set when all
the DIP switches are OFF
or when DIP switch 1 is
ON. For MR Protocol,
Address 0 is set when all
DIP switches are OFF.
Table 3-2 details the addresses set by the binary DIP switches 1-6.
Table 3-1 DIP Switch Settings
Position #
Description
Setting
1
Binary Bit 0
Addressing, OFF=1, ON=1
2
Binary Bit 1
Addressing, OFF=1, ON=1
3
Binary Bit 2
Addressing, OFF=1, ON=1
4
Binary Bit 3
Addressing, OFF=1, ON=1
5
Binary Bit 4
Addressing, OFF=1, ON=1
6
Communication
Mode
Always ON for MODBUS ASCII
communication
7
Communication
Mode
Always OFF for proper operation
8
Communication
Mode
Default OFF - when ON, records
are not saved to buffer
Table 3-2 DIP Switch Addresses
DIP SWITCHES
1 2 3 4 5 6
ADDRESS
1 0 0 0 0 0
1
0 1 0 0 0 0
2
1 1 0 0 0 0
3
0 0 1 0 0 0
4
1 0 1 0 0 0
5
0 1 10 0 0
6
1 1 1 0 0 0
7
Summary of Contents for REMOTE ActiveCount
Page 1: ......
Page 2: ...00...
Page 6: ......
Page 10: ...Lighthouse REMOTE ActiveCount Installation Guide ii 248083413 1 Rev 4...
Page 12: ...Lighthouse REMOTE ActiveCount Installation Guide 1 2 248083413 1 Rev 4...
Page 32: ...Lighthouse REMOTE ActiveCount Installation Guide A 2 248083413 1 Rev 4...
Page 50: ...Lighthouse REMOTE ActiveCount Installation Guide B 18 248083413 1 Rev 4...
Page 53: ...00...
Page 54: ......