BPM 1000i
Blood Pressure Monitor
Error Messages
ERR 1: Cuff tube kink, blocked or inflation is too slow (<2 mmHg/sec).
ERR 2: Cuff leakage or inflation is too fast (>10 mmHg/sec).
ERR 3: Detected arm movement.
ERR 4: No pulse is detected.
ERR 5: Cannot figure out diastolic or systolic pressure or it is out of legal range.
ERR 6: Cuff pressure is too high (>300mmHg).
ERR 7: The cuff pressure cannot return to 0 mmHg when valve is open.
17
Display System
Liquid Crystal Display
Measuring Method
Oscillometric method
Power Source
6VDC; four "AA" type batteries (1.5V)
Measuring Range
40 ~ 250mmHg
40 ~ 200 pulse/minute
Accuracy
±
3mmHg
±
5% pulse rate
Inflation
Electrical micro-rolling pump
Pressure Exhaust
solenoid valve
Memory
1.
6 Users with 30 records per user.
2.
External USB flash drive.
Display
year/date/time/pressure/pulse/memory
Low Voltage Indication
Yes
Auto Power off
150s without any operation
Battery Life
around 1000 operations (new alkaline batteries)
Operation Environment
+10°C ~ +40°C (+50°F ~ +104°F); RH ≤ 90%
Storage/Transportation
Environment
-20°C ~ +50°C (-4°F ~ +122°F); RH ≤ 95%
Device lifetime
three years with 25 measurements per day
Dimension
around 162 (W) x 120 (D) x 62 (H) mm
Weight
around 400g (including batteries)
Product Specifications
Warranty
This device is covered by a two year guarantee from the date of purchase.
The guarantee is valid only on presentation of the original sales receipt.
The warranty covers both the instrument and the cuff (batteries are not
included). Abnormal use or any customer disassembly of the unit will
void the warranty.