Operation modes
Introduction
The PowerEase
®
90W Power Supply is a microprocessor-controlled power supply
designed to meet most electrophoresis needs in a single, easy to use unit. The power
supply offers two modes: Constant Voltage and Constant Current. Use the Constant
Voltage or Constant Current Mode for applications that require only one specific voltage
limit or current limit during the entire duration of electrophoresis.
Constant operation
protocol
Instructions for operating the PowerEase
®
90W Power Supply in the
Constant Mode
are
provided in this section. The
Constant Voltage
or
Constant Current Mode
allows you to
specify a voltage limit or current limit to be used during the entire duration of
electrophoresis. Review the guidelines provided in this manual before starting
electrophoresis using the PowerEase
®
90W Power Supply.
A basic
Constant Voltage
or
Constant Current Mode
operating procedure of the
PowerEase
®
90W Power Supply is provided below. We recommend reading the
guidelines provided in this manual for best results before starting an operation.
1.
Use the power switch on the rear side of the instrument to turn on the PowerEase
®
90W Power Supply.
The display screen will illuminate.
The screen displays the last set
V
(voltage) in large bold type on the left side of the
display and
TIME
and
AMP
in small type on the right side of the display.
2.
Press the
Constant
key to select either
Constant Voltage
or
Constant Current
from
the display screen.
The screen displays the selected operating mode in large bold type on the left of the
display.
3.
Use the
∧
∨
arrow keys to set the selected operation [voltage (VOLT) or current
(AMP)] to the appropriate values.
4.
Use the
Mode
key to select
TIME
and use
∧
∨
arrow key to set the time
(hours/minutes) to specify the duration of the electrophoresis.
5.
Press
Start/Pause
key to start electrophoresis. The Red LED illuminates when the
electrophoresis is running.
6.
Press the
Start/Pause
key again to temporarily interrupt power to ongoing
electrophoresis without terminating the operation along with LED flashing.
7.
Press
Start/Pause
key to restart electrophoresis.
8.
Press
Stop
key to stop the electrophoresis.
Note:
If the parameters need to be changed during the run you must press
Start/Pause
, enter the
changes, and then press
Start/Pause
once again to restart your operation.
Note
: After stopping and restarting an operation, the timer resets to the selected time
Limiting parameter settings
Introduction
The PowerEase
®
90W Power Supply is capable of operating at limiting Voltage, or
limiting Current. We recommend operating the PowerEase
®
90W Power Supply at
limiting voltage for most applications. See below for more details.
8
PowerEase
®
90W Power Supply User Guide