Page 11
Instruction Manual V1.3
This is used as a power switch as well as an emergency cut off switch. To
use the unit, first turn this breaker on. If by some reason the internal safety
features do not work, this circuit breaker shut off the AC power from a
user’s
power line.
1.6.3 Output Terminal/Direct Stripline
1) Direct Stripline:
If the output is not 50ohm, there a special made transmission line is used to
connect a load to the unit. The stripline is designed
for the customer’s
specific load impedance to carry a high current without destroying the current
pulse shape (avoiding the reflective wave caused by the impedance miss-
matching). One side of the strip line is just open to the load to connect by
soldering/screws.
See 2.3 section for detail.
2) H.V Switch Output
This is the H.V output for a 50ohm BNC. Customer can connect this output to
the customer’s 50ohm load.
3) H.V Switch Input
This is the H.V switch input for a 50ohm BNC. Customer can use this input to
switch customer’s H.V output. Usually this input is connected to H.V Power
Supply Output with a 50ohm H.V coaxial cable supplied.
4) H.V Power Supply Output
This is the H.V output from the H.V power supply built in the unit. Usually this
output is connected to H.V Switch Input with a 50ohm H.V coaxial cable
supplied.
1.6.4 Water Inlet for 25KW model
Water inlet is ¼” NPT connector. Required water flow is 5 liter/minute
minimum and the water pressure is 3kg/cm
2
minimum required.