background image

Fuses

Check the continuity of the fuse. With power OFF, 

pull the fuse out of the fuse block. With the ohm-

meter on the R X 1 scale, measure resistance. A 

reading of infinity indicates a blown fuse and must 

be replaced. Replace fuse with same type, voltage 

and amp rating.

Indicator Lights

If defective, replace all indicator pilot lights with 

same type.

Magnetic Contactor

Contacts

 - Check the contacts for severely burnt 

or welded contacts.  The contactor arm should 

move freely.

Coil

 - Measure the coil by disconnecting one of 

the coil leads.  Measure the coil resistance by set-

ting the ohmmeter on the R X 1 scale.  A defective 

coil will read zero or infinity, indicating a short or 

opened coil respectively. Replace defective con-

tactor with same type.

NOTE: 

Readings may vary depending on accuracy 

of the measuring device.

Motor Protective Switch

Test by inserting a paper clip or other small device 

into the test hole and push to the left. The relay 

should trip. 

Alarm Horn

Moving the alarm test/normal/silence switch to the 

test position or activating the alarm float should 

turn on the alarm horn. If the horn does not sound 

replace with horn of same type.

Alarm Light

Moving the alarm test/normal/silence switch to the 

test position or activating the alarm float should 

turn on the alarm light. If the light does not activate, 

replace with bulb of same type.

Float Controls

Check the floats during their entire range of opera

-

tion. Clean, adjust, or replace floats.

Checking the float resistance

 - The float resis

-

tance can be measured to determine if the float 

is operating correctly or is defective.  Use the fol-

lowing procedure to measure the float resistance:

 

1.

   Isolate the float by disconnecting one or both 

of the float leads from the float terminals.

2.

   Place one ohmmeter lead on one of the float 

wires, and the other ohmmeter lead on the 

other float wire.

3.

   Place the ohmmeter dial to read ohms and 

place on the R X 1 scale. With the float in the 

“off”  position,  the  scale  should  read  infinity 

(high resistance). Replace the float if you do 

not get this reading. With the float in the ON 

position,  the  scale  should  read  nearly  zero 

(very low resistance).  Replace the float if you 

do not get this reading.

NOTE: 

Readings may vary depending on the 

length of wire and accuracy of the measuring 

device.

Troubleshooting

Installation of Floats

Standard three phase simplex panels are designed 

to operate with three floats. These floats operate the 

pump stop, pump start, and high level alarm functions.

NOTE: Options ordered may affect the number 

of  floats  and  their  functions.    Please  reference 

the schematic provided with the control panel for 

proper installation.

CAUTION: 

If control switch cables are not wired and 

mounted in the correct order, the pump system will not 

function properly.

WARNING:

 Turn off all power before installing floats in 

pump chamber. Failure to do so could result in serious 

or fatal electrical shock.

1. 

 Use float label kit to identify and label cables on 

both the float and stripped ends (stop, start, alarm, 

etc.). See schematic for float options.

2. 

 Determine your normal operating level, as illustrated 

in 

Figure 1

.  

3. 

Mount  float  switches  at  appropriate  levels  as  il

-

lustrated in 

Figures 2-4

. Be sure that floats have 

free range of motion without touching each other 

or other equipment in the basin.

If using the mounting clamp; follow steps 4-6.

4. 

Place the cord into the clamp as shown in

 Figure 

2

.

5. 

Locate the clamp at the desired activation level and 

secure the clamp to the discharge pipe as shown 

in 

Figure 2

.  

 NOTE:

 Do not install cord under hose clamp.

6. 

 Tighten the hose clamp using a screwdriver. Over 

tightening may result in damage to the plastic 

clamp. Make sure the float cable is not allowed to 

touch the excess hose clamp band during operation.

 NOTE:

 All hose clamp components are made 

of 18-8 stainless steel material. See your SJE-

Rhombus

®

 supplier for replacements.

Installation Instructions

FIGURE 4: 

Float with cable weight

FIGURE 1:

 Three float simplex - 

pump down installation 

FIGURE 2

: Mounting clamp detail

FIGURE 3: 

Internally weighted float

- 2 -

- 5 -

7227000B

7227000B

Reviews: