Section 4. Software overview
Calibration
User calibration settings are configured under
Options > Calibration
.
Zero
gives the date of the most recent zero calibration. See
page 6-1 for details on the Zero button.
N
2
O Span
presents information about the previous span setting, including the date
that the span was set (
Last
), the span value (
Weight
; the multiplier that is set when
you span the N
2
O measurement), and the N
2
O
Concentration
(ppb) entered in the
previous span. Do not change the settings unless you are certain that your actions
will improve the performance of the instrument. See
Restore Span/Zero
allows you select a previous zero and span setting.
Factory Reset
reloads the factory original zero and span calibration parameters.
4-10
LI-7820 N₂O/H₂O Gas Analyzer