Section 6. Maintenance
Setting the CO₂ and H₂O zeroes
Select a quality zero gas that is free of volatile organic compounds (VOCs). The gas
should be 0 ppm CO
2
in air; the calibration will only be as good as the gas.
Caution:
Do not use nitrogen-balanced gas. Use air-balanced gas for both the zero
and span. Do not power on the instrument with a 0 ppm CO
2
gas flowing
through the analyzer or the analyzer will not initialize properly.
1
Power on the instrument and let it warm up.
2
Plumb a gas to the air inlet as shown in
Set the pressure between ambient and 35 kPa (5 PSI) above ambient.
Observe the measurements on the home screen until they have stabilized. The slope
will be near zero.
3
Zero the analyzer.
Under
Options > Calibration
, click
Zero
. The software will indicate that a zero cal-
ibration has started and it will not report concentrations until the zero is set. This
usually takes about 5 minutes.
6-2
LI-7815 CO₂/H₂O Gas Analyzer
Summary of Contents for LI-7815
Page 1: ...LI 7815 CO2 H2O Trace Gas Analyzer Instruction Manual...
Page 2: ......
Page 10: ...viii...
Page 30: ...Section 2 Getting started 2 14 LI 7815 CO H O Gas Analyzer...
Page 38: ...Section 3 Applications of the LI 7815 3 8 LI 7815 CO H O Gas Analyzer...
Page 42: ...Section 4 Software overview 4 4 LI 7815 CO H O Gas Analyzer...
Page 54: ...Section 4 Software overview 4 16 LI 7815 CO H O Gas Analyzer...
Page 62: ...Section 5 Troubleshooting 5 8 LI 7815 CO H O Gas Analyzer...
Page 70: ...Section 6 Maintenance 6 8 LI 7815 CO H O Gas Analyzer...
Page 88: ...Appendix B Additional safety information B 6 LI 7815 CO H O Gas Analyzer...
Page 92: ...Appendix C Specifications C 4 LI 7815 CO H O Gas Analyzer...
Page 97: ......