![LEYBOLD SCROLLVAC plus Service Manual Download Page 7](http://html1.mh-extra.com/html/leybold/scrollvac-plus/scrollvac-plus_service-manual_1890850007.webp)
7
300668736_002_C0 - 11/2017 - © Leybold
6. Starting from one end, remove the tip seal from the fixed scroll. Repeat
this operation for the orbiting scroll (Figure 5, item 3). Remove the fixed
scroll O-ring (Figure 5, item 4). Dispose of all used items that have been
removed following this operation.
Note: Avoid damaging the fixed scroll O-ring groove when removing the
O-ring.
7. Using a clean dry lint-free cloth and IPA, remove all dust and process
debris from the scroll forms and grooves. Heavier process debris can be
removed with Green Scotch Brite
®
only.
8. Taking a new tip seal, fit it using the following technique in both the orbit-
ing scroll and the fixed scroll:
Figure 5 - Replacing the tip seals
1 Housing assembly
2 Orbiting scroll
3 Tip seal
4 Fixed scroll O-ring
5 Fixed scroll
6 Fixed scroll fixing screws
7 Fan-cowl
8 Alignment dowel
1
3
4
3
5
6
7
8
2