GA 12.107/5.02 - 12/97
Description
6
1.3 Technical Specifications
Helium filling pressure at room temp.
Operation at 50 Hz
bar
16
-1
psig 218 to 232
Operation at 60 Hz
bar
14
-1
psig 189 to 203
Operating pressure in the high-pressure
circuit at 50 Hz
bar
22±1
psig 305 to 334
and with connection of an
RGD
5/100-1
Permissible ambient temperature
2)
°
C
5 – 40
°
F 40 to 105
Permissible relative humidity, max.
%
90
Coolant inlet temperature
2)
°
C
5 – 25
°
F
40 to 80
Permissible coolant connection
bar
8
pressure
psig
116
Coolant connection, hose nipples Nom.diam
12
Coolant consumption at 25
°
C
(approx. 80
°
F) inlet temperature
l/min
5
gal/min
1.3
Max. coolant flow rate
l/min
10
gal/min
2.6
Mains power supply, 50Hz, 3-ph.,
each±5%
V
400, 415
Mains power supply, 60Hz, 3-ph,
each±5%
V
460, 480
Delivered from the factory wired for
V
415
Electrical power drawn at 50 Hz
kW
6.1 – 6.9
Electrical power drawn at 60 Hz
kW
6.5 – 7.2
Operating current at 50 Hz
A 9.8 – 11.1
Operating current at 60 Hz
A 8.6 – 10.2
Mains cord connection plug
3)
Type
4
Control voltage for the remote control V DC
24
Connection voltage for the cold head
V AC
240
Helium connectors: self-sealing couplings
4)
1/2“ m
Noise level
5)
, at a distance of 1 m
dB(A)
78 - 80
Dimensions (W x H x D) mm 450x445x445
Weight
kg
105
lbs
232
1) --
2) Operation at lower temperatures is possible if the appropriate
measures are taken. Please consult the manufacturer for details.
3) Type 4: four conductor ferrules 4,0 mm
2
x 18 mm long (0.7 inch)
with designations L1/L2/L3/PE
4) f=female / inside threads; m=male / outside threads „Aeroquip
series 5400“ or compatible designs.
5) As per DIN 45 635, Part 1, corresponding to ISO 6081-1986
1.4 Order Information
Cat.-No.
Compressor unit, RW 6000-1
891 44
Adsorber, ARW 2
891 84
Flexlines
One pair each with self-sealing couplings
at both ends
FL 4.5 (
1
/
2
“ f
4)
,
1
/
2
“ f), 4.5 m long (15 feet)
892 87
FL 9.0 (
1
/
2
“ f,
1
/
2
“ f), 9 m long (29.5 feet)
892 88
FL 2.0 ELBOW (
1
/
2
“ m /
1
/
2
“ f), 2 m long
893 66
FL 13.5-SF (
1
/
2
“ f,
1
/
2
“ f), 13.5 m long (45 feet)
893 67
FL 11.3 (
1
/
2
“ f,
1
/
2
“ f), 11.3 m long (37 feet)
893 68
Installation accessories
with self-sealing couplings, Helium filling
pressure at room temperature (16
-1
) bar (232 psig)
90
°
elbows,
1
/
2
“ m /
1
/
2
“ f
891 73
Insulating adapter,
1
/
2
“ m /
1
/
2
“ f
891 76
Line coupler,
1
/
2
“ m /
1
/
2
“ m
891 71
CRYO SERVICE SET, case with tools for
maintenance work at all cold heads and
compressor units
890 93