6.11.4 ECODRY plus Protocol
Table 15 Read commands
Command Description
Read: Run second
Reads the run seconds of the pump
Slave Address
0x01
Function Code
0x03
Start Address Upper
0x00
Start Address Lower
0x80
Data Quantity Upper
0x00
Data Quantity Lower
0x02
CRC-16 Upper
-
CRC-16 Lower
-
Table 16 Write Commands
Command
Description
Slave Address
Function
Code
Start Address
Data quantity
Number
of Bytes
Starting Data
CRC-16
Upper
Lower
Upper
Lower
Upper
Lower
Upper
Lower
Write:
START
STARTS the
pump via
MODBUS
0x01
0x10
0x00
0x50
0x00
0x01
0x02
0x00
0x02
TBD
TBD
Write:
STOP
STOPS the
pump via
MODBUS
0x01
0x10
0x00
0x50
0x00
0x01
0x02
0x00
0x01
TBD
TBD
Operation