4. 

Operation 

4.1 

Switching ON and Switching OFF 

  Switch the instrument on through its main switch. 

 

After switching on, the GRAPHIX controller will run the following:

 

 

  Display of the start screen with the version number 

  Re-establishing of the most recently setup parameters  

  Identification of the connected measuring instruments 

  Enabling of the measurement mode with measurement display mode Normal, 

Big or Chart (depending on the most recent setting) 

 

  Switch the instrument off through its main switch. 

 

 

CAUTION: Waiting Time 

Wait for at least five seconds before switching the instrument on again. 

 

4.2 

Display 

The GRAPHIX controller makes different display modes possible. You may select between 
the following display modes: 

 

Normal 

Default display mode, which provides all important information on the connected sensors. 

 

Chart 

Graphic  display  of  the  pressure  history  of  the  connected  sensors.  Besides  the  pressure 
history, also the measured values for the individual channels are displayed. 

 

Big 

The display is limited to displaying the measured values of the connected sensors using a 
large font. 

 

Speedo 

Pressure  reading  for  the  connected  sensor  of  one  selected  channel  is  in  the  shape  of  a 
speedometer.  The  mantissa  is  displayed  as  a  round  progress,  whereas  the  exponent  is 
displayed centrally. 

 

Leak Test 

Display mode for the Leak Test function. Besides the leak rate, current pressure, total time 
and remaining time are displayed. 

 

The selection is made using the channel menu button 

  or the parameter Display Mode in 

the parameter group Display of the main menu. 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for 230680V01

Page 1: ...Catalog No 230680V01 230681V01 230682V01 GRAPHIXONE GRAPHIXTWO GRAPHIXTHREE Vacuum Gauge Controller Operation Manual GA300553102_002_C2...

Page 2: ......

Page 3: ...Symbols 10 4 4 1 Symbols for the Controls 10 4 4 2 Symbols for the Language Selection 11 4 4 3 Status and Warning Symbols 11 4 5 Measuring Mode 12 5 Maintenance and Service 13 5 1 Maintenance 13 5 2 C...

Page 4: ...HIX controller is a display and operating unit with a graphic user interface for sensors with an analogue or digital interface of the company Leybold GmbH or from other manufacturers Depending on the...

Page 5: ...sume any responsibility or warranty in case the operator or third persons do not observe the information given in this document do not use the product as intended modify the product in any way convers...

Page 6: ...h electricity NOTICE General information pointing to further information respectively reference sections 2 3 Basic Safety Information During all work like installation maintenance and repair activitie...

Page 7: ...measurement channel 2 for GRAPHIX TWO and THREE only for sensors CH2 A and CH2 B E1 E2 Connection of measurement channel 3 for GRAPHIX THREE only for sensors CH3 A and CH3 B F Relay output connection...

Page 8: ...sible You may select between the following display modes Normal Default display mode which provides all important information on the connected sensors Chart Graphic display of the pressure history of...

Page 9: ...e functions In this way you may configure your GRAPHIX controller and also utilise further functions offered by the instrument Channel Menu Buttons Briefly touching the desired channel invokes the cha...

Page 10: ...Enabling display mode Leak Test Speedo Channel 1 Enabling display mode Speedo for channel 1 Speedo Channel 2 Enabling display mode Speedo for channel 2 Speedo Channel 3 Enabling display mode Speedo f...

Page 11: ...ge RU Russian Polish Select menu language PL Polish Turkish Select menu language TR Turkish Table 4 Symbols for language selection 4 4 3 Status and Warning Symbols Symbol Designation Explanation Statu...

Page 12: ...in specified measuring range Measuring value and additional warning symbol in the display mode Normal and Speedo or Measuring value displayed by a yellow font in the display modes Chart Big and Leak...

Page 13: ...n the USB stick in the folder RECOVERY o After the reset process has been completed the instrument will automatically restart with the default parameters o The GRAPHIX controller will now be ready for...

Page 14: ...Setup o The question Are you sure is displayed to ensure that you really want to run the reset When not wanting to run the reset tap the button to cancel the reset process o The main menu is then disp...

Page 15: ...yed again To run the update tap the button Start update o On the memory at the USB interface is installed the folder RECOVERY o Configuration data are stored in a file with the format rescue txt on th...

Page 16: ...measurement channel Acknowledge this message by selecting and terminating the channel menu at the corresponding channel Analog sensor signal out of range too high Error H 1 4 The measurement signal f...

Page 17: ...memory stick Error usb file closing none 3 File on USB stick cannot be stored Check the attached USB memory stick Error usb write none 4 USB stick not recordably Check the attached USB memory stick Er...

Page 18: ...n The setup value is saved To exit tap the button o The parameter selection display is displayed again 6 3 Help in Case of Malfunctions If the malfunction persists even after having exchanged the sens...

Page 19: ...ry room During storage the following ambient conditions need to be maintained Ambient temperature 20 60 C Humidity of the air As low as possible Preferably in a sealed plastic bag with desiccant 7 3 W...