background image

levr

 

LIM-05 user manual 

Dated 25/5/2017 

http://www.levr.pl/

 

Page 

13

 of 

20

   

If  the  LIM-05  instrument  is  unable  to  automatically  “locate”  the  end  of  the  wire  for 
any reason, conduct the calibration described in section 0 of this manual. 
METHOD 2 
The specific procedure is described in section 8.1.  
 
Note: A new reference run should be created and recorded after changing the physical 
length of the sensor wire. (See section 6.4 i 0). 

6.2  “Napi

ę

cie progowe Up=” (Voltage threshold Up=) entry 

Input signals whose voltage amplitude is within the ±Up range are not analysed. This 
limitation  can  be  used  to  eliminate  some  of the  low-amplitude  interference  appearing 
in the alarm system of the district heating pipeline. The ±Up voltage range can be set 
separately for each measurement channel. 

6.3  “Liczba kanałów=” (Number of channels =) entry 

The  instrument  can  monitor  up  to  four  sensor  loops.  The  entered  number  (1–4) 
specifies  the  number  of  monitored  sections  of  the  district  heating  pipeline  with  the 
alarm system. (See section 0). If the number of connected loops is smaller, they have 
to  be  connected  to  subsequent  measurement  ports  of  the  instrument  starting  from  the 
outermost left port (channel 1). 

6.4  “Przebieg odniesienia” (Reference run) entry 

Activating  this  function  will  force  a  specific  measurement  cycle  in  the  selected 
measurement  channel.  The  instrument  will  calculate  the  average  value  of  measured 
results i, which will be recorded as the reference run. (See section 0). 

6.5  “Czas martwy=” (Dead time=) entry 

The automatic measurement cycle of the fault locator consists of measurements carried 
out for every section of the district heating pipeline. The measurements are carried out 
sequentially.  Between  the  measurement  cycles,  there  is  a  gap  referred  to  as  the  dead 
time. The dead time can be adjusted in the range of 1–255 min. (See section 0). 

6.6  “VOP=” entry 

This entry enables setting the VOP coefficient for all measurement channels in use. It 
may be a fraction in the range of 0.3–1.0 or a value expressed as percentage from 30 to 
100%. (See section 0). 
The factory setting is: VOP = 0.92. In practice, the coefficient is set as 0.84–0.93 for 
measurements of district heating pipelines.  

6.7  “Parametry impulsu>” (Pulse parameters>) entry 

The  amplitude  in  [V]  and  duration  in  [ns]  of  the  measurement  pulse  can  be  set  for 
every measurement channel. Both of these values are selected depending on the length 
of the monitored district heating pipeline (length of the monitored sensor wire).  
1-czas trwania =” (1 – duration =) – duration of the measurement pulse is changed 
with  keys  marked  with  “

”  (increase)  and  “

”(decrease)  symbols.  The  changes  are 

implemented in a stepwise fashion, at 12.5-ns intervals within a range of 12.5–75 ns.  
2-amplituda impulsu =” (2 – pulse amplitude) – the amplitude of the measurement 
pulse  can  be  changed  within  a  range  of  2.3–7.8 V  (4.8–14.6 V  for  amplitude  of  the 
voltage  of  the  measurement  pulse  generator)  using  keys  marked  with  “

”  (increase) 

and “

”(decrease) symbols. 

 

Summary of Contents for LIM-05

Page 1: ...LIM 05 pulse fault locator user manual Software version 17 05 USER MANUAL EZUP LEVR s c 03 193 Warsaw ul Krzy wki 5 tel fax 48 22 639 85 66 Ver 17 12...

Page 2: ...0 FN key 9 4 11 ENT key 10 4 12 ESC key 10 4 13 key 10 4 14 and keys 10 5 Main menu 10 5 1 Kana Channel option 10 5 2 Start Stop option 10 5 3 Kalibracja Calibration screen 10 5 4 Historia History opt...

Page 3: ...16 Calibration procedure Kalibracja Calibration option D ugo p tli Loop length entry 16 Channel number entry 18 Initiate or stop automatic measurements 18 Creating and recording the reference run 18...

Page 4: ...t calculates the distance between the measurement station and the point of reflection i e location of the fault The method of fault location in preinsulated district heating pipelines with a detector...

Page 5: ...epicts the actual distance in m Figures 1 and 2 depict the method of presentation of measurement information Fig 1 Depiction of the reflected measurement pulse on the detector screen in case of a leak...

Page 6: ...ength of the section of the alarm wire between these cursors x default axis x y default axis y Up Up voltage thresholds for initiation of automatic fault detection and indication the dead area is 2Up...

Page 7: ...he operating conditions return to normal 3 1 Connection of the locator to the alarm system of the district heating pipeline The locator is connected to the carrier pipe and copper sensor wire with a c...

Page 8: ...nment and the speed of light propagation in vacuum For district heating pipelines with a copper sensor wire this coefficient adopts values of the 0 91 0 93 range 4 2 2 zoomH key Press the 2 zoomH key...

Page 9: ...of the screen after pressing the 6 K2 key 4 7 7 PING key Activating the PING function will force a measurement cycle in the selected measurement channel and display the measured run Depending on the...

Page 10: ...ia Reference run 8 Info 5 1 Kana Channel option This option can be used to display measurement information about one of the monitored sections of the district heating pipeline To select the channel en...

Page 11: ...iowy Input run option enables presentation of the collected digital data in the form of a chart These data refer to the last measurement When the input run is displayed the screen will display a descr...

Page 12: ...ion method also ensures that the image corresponds to reality For instance if a fault occurs near the end of the network it will be represented near the end of the x axis and not for instance in the c...

Page 13: ...ge value of measured results i which will be recorded as the reference run See section 0 6 5 Czas martwy Dead time entry The automatic measurement cycle of the fault locator consists of measurements c...

Page 14: ...nitial stages of its development The shape of the reflected pulse will be represented within the range of voltage thresholds determining the measuring range Extending the voltage measuring range will...

Page 15: ...ce of this message means that there is no communication between the computer and the locator In such a situation make sure that the devices are correctly connected and check the status of the selected...

Page 16: ...ed only after the locator screen displays the run Calibration procedure Kalibracja Calibration option D ugo p tli Loop length entry Note Locator calibration should be carried out 30 minutes after the...

Page 17: ...he section of the x axis visible on the chart does not reflect the actual length of the alarm wire In section 3 increased length of the line was specified precisely in order to ensure that the image o...

Page 18: ...on of automatic measurements is indicated by the AUTO description It will appear in the bottom right corner of the screen Stopping automatic measurements will cause the description to disappear The se...

Page 19: ...1 10 0 M Ma ai in nt te en na an nc ce e o of f t th he e L LI IM M 0 05 5 i in ns st tr ru um me en nt t A clean dry cloth should be used to remove dust from the housing of the instrument Other dirt...

Page 20: ...range 2047 mV Maximum measurement sensitivity 1 mV pixel Sensitivity threshold voltage range 1 1024 mV Dead time range 1 255 min Methods of communication with external devices Alarm output volt free r...

Reviews: