2.3 General Environmental Conditions
Usage
Indoor
Altitude
Up to 2000
m
Operating ambient temperature
0 to 40 °
C
Storage ambient temperature
(Extremes for Transportation)
-20 to 80 °
C
-20 to 70 °
C
for single-use pump head
Operating humidity range (relative humidity)
15
– 95
%
(non-condensing)
Storage humidity range (relative humidity)
(Extremes for Transportation)
15
– 95
%
(non-condensing)
Normal storage conditions
Ambient temp.: 20 to 30 °
C
Relative humidity: 50
%
(non-condensing)
DC supply fluctuations
10% of nominal voltage
Transient over-voltages typically present on
the mains supply
Surge immunity according to EN 61000-4-5
(tested together with certified AC/DC power supply)
Pollution degree
2
Table 5: Environmental conditions for flow control system
2.4 Pressure-Flow Curves
Figure 8: Pressure/flow rate curves
(Typical curves measured with DCP-30.1 pump head.)
0
2
4
6
8
10
12
14
16
0.0
0.5
1.0
1.5
2.0
0.0
0.2
0.4
0.6
0.8
1.0
0
1
2
3
4
5
6
7
8
[psi]
[gallons / min]
[bar]
[liters / min]
6000 rpm
13000 rpm
12000 rpm
11000 rpm
10000 rpm
8000 rpm
Specific gravity = 1g/cm
3
Viscosity = 0.7cP
Liquid Temp.: 40
C
9000 rpm
7000 rpm