Two Fieldbus connectors (IN and OUT) allow to setup arrays of multiple flow controllers. Therefore, blending
configurations as shown in
Figure 6: Flowcontrol array for blending applications with EasyConnect models
2.2.3 System Configurations for OEM Models
The
OEM
model, with one integrated driver connector only, is designed for compact integration into an
equipment. The flow can precisely be set via an analog input. Various digital inputs and outputs allow
controlling and monitoring of the system. A
RS485 (Modbus)
enables communication with a user panel or a
PC with the
Levitronix
®
Service Software
. Hence parameterization, firmware updates and failure analysis are
possible. The
RS485
can also be used as a fieldbus to implement more intelligent concepts of pump control.
Figure 7: System configuration of OEM model
Integrated Flow Control
Driver
EasyConnect Model
P
o
w
e
r
S
u
p
p
ly
P
L
C
F
ie
ld
b
u
s
O
U
T
F
ie
ld
b
u
s
IN
Single-Use
Pump Head
Fluid 1 Out
Fluid 1 In
IP
S
C
a
b
le
S
ig
n
a
l
LEVIFLOW
®
Flow Sensors
24 VDC
Supply
IP
S
C
a
b
le
P
o
w
e
r
Remote Control
over RS485
IP
S
C
a
b
le
S
ig
n
a
l
IPS Cable Signal
Integrated Flow Control
Driver
EasyConnect Model
P
o
w
e
r
S
u
p
p
ly
P
L
C
F
ie
ld
b
u
s
O
U
T
F
ie
ld
b
u
s
IN
Single-Use
Pump Head
Fluid 2 Out
Fluid 2 In
IP
S
C
a
b
le
S
ig
n
a
l
LEVIFLOW
®
Flow Sensors
24 VDC
Supply
IP
S
C
a
b
le
P
o
w
e
r
Blended
Fluid
Fieldbus
Termination
Connector
Integrated Flow Control
Driver OEM Model
PLC Interface:
- Analog and Digital IO
s
- Remote Control
AC/DC
Power
Supply
(24VDC, DIN-Rail)
AC Input
DC Output
PC Connection:
- System configuration
- Firmware update
- Monitoring
Open
Wires
Screw type connector for
open-wires connection
included in cable package.
Fieldbus
RS485
of Driver
Integrated
Driver Cable
Si
n
g
le
-U
s
e
Pu
m
p
H
e
a
d
Bayonet Socket
with Locking Pin
USB/RS485 Cable
IPS Cable Signal
Flow Out
L
EVI
F
L
O
W
®
F
lo
w
Se
n
s
o
rs
User Panel
24VDC
9
9
10
2b
2b
1
1
10a
10a
10a
13
4
5
4
5
2c
1
7a
8
3
4
5
12
10a
15
10b
Pin Designation
LUI-B.1 Cable 10b:
1: RS485+
2: RS485-
3: GND
4: NC
5: NC
6: +24VDC