Instructions for Use of LEVIFLOW® LFS-SU Flowsensors
www.levitronix.com
PL-4509-01, Rev01, DCO# 16-191
1/2
Technical Support:
Levitronix
Technical Service Department
Technoparkstr. 1, CH-8005 Zurich, Switzerland
Phone for US:
888-569 07 18
Phone for outside US:
+1 888-569 07 18
E-Mail:
1 Description and Preparation
1.1 Description
The
LEVIFLOW®
LFS-SU
flowsensors are
designed for non-invasive flow measurements in
single-use applications of the Life Science
industries. The wetted materials of the sensors are
made out of biocompatible polypropylene (FDA,
USP-VI, BSE/TSE and Animal free), which can be
gamma sterilized.
Figure 1: Single-use flowsensor (example LFS-06SU)
Before using the flowsensor make yourself familiar
with the following instructions.
1.2 Inspection Prior to Use
The flowsensor should be inspected prior to use for
any damage. Do not use the flowsensor if any
damage is found. Contact
Levitronix
®
regarding
return of any suspected flowsensor.
1.3 Traceability for Troubleshooting
For full traceability of the flowsensor the serial
number located on the sensor body (see
Figure 1
)
shall be used. The serial number is also printed on
the biocompatibility declaration and the calibration
sheet delivered with each sensor.
1.4 Levitronix
®
Configuration Software
For sensor parameter setting and calibration
parameter transfer to the converter (not necessary
for sensors with a parameter chip) the
Levitronix
®
Configuration Software
with version
V1.30
or
newer is needed. For details regarding installation
of software and establishing RS485 connection
consult the
LEVIFLOW
®
Configuration Software
manual (see Doc#
PL-4501-00)
.
2 General Warnings and Cautions
CAUTION
Gamma Sterilization
The LFS-SU flowsensors have been tested to be
robust against gamma radiation with a dose up to
40 kGy.
Handle the flowsensors with care, especially after
gamma radiation with doses above 25 kGy, since the
radiation exposure decreases the flexibility of the
flowsensor material.
3 Procedure for Installation
3.1 Mechanical/Electrical Installation
Connect the sensor to the converter together with
the relevant adaptor cables. Take specific care
about the flow direction indicated with the arrow on
the flow sensor (see
Figure 1
). For details about
mechanical and electrical installation and design of
an adequate hydraulic circuit, consult the user
manual (Doc#
PL-4509-00
).
3.2 Setting of Sensor Type Parameters
The sensor type (or size) specific parameters can
be set or updated on the converter at the first
installation or whenever the sensor type is
changed.
Note: Sensors which are labeled with “Type: PC”
on the calibration tag have an integrated
parameter chip and parameters are transferred
automatically to the converter.
Start the
LEVIFLOW
®
Configuration Software
and
open the “
Parameter Settings
” menu (
).
Make sure that “
Sensor Size”
, “
Full Scale”
, “
Kinem.
Viscosity”
and “
Liquid Temperature
“ are set to the
values designated for the used sensor type.
table shows the standard values. For other
configuration consult the specific sensor
specification.
Sensor
Type
Full Scale
[L/min]
Kinem.
Viscosity
[cSt]
Liquid
Temperature
[°C]
LFS-03SU
0.800
1
20
LFS-06SU
8.000
1
20
LFS-10SU
20.000
1
20
LFS-15SU
50.000
1
20
LFS-20SU
80.000
1
20
Table 1: Standard sensor type parameters
Calibration
Parameter
Tag
Measurement Path
Connector
Box
Label with Serie # and
Flow Direction Arrow
Inlet
Outlet
Fixation Hole