Shenzhen LEPU Intelligent Medical Equipment Co., Ltd.

Form No.: LP-QR-RD-016 REV.01

Page 2of 7

1.4 Install or replace batteries

The first step after unpacking is to install the battery. The battery holder is on the back of the

infrared forehead thermometer. The battery installation steps are as follows:

(1) Open the battery cover.

(2) Insert the battery and keep +, - poles of the battery in line with +, - poles of the battery holder.

(3) Close the battery cover.

Note:

Please use 2 AAA alkaline batteries.

Please use batteries if the same type, do not mix used and new batteries.

Remove the battery when the thermometer will not be used for a long time.
The symbol “

” appearing on the screen indicates that the battery runs out; please replace the

battery before measuring.

1.5 Starting up

Press the "ON/OFF" button, then the backlight of the display comes on, the LCD is shown

in full screen and displays the latest group of memory values, the equipment enters the waiting

state for measurement; at this time, the backlight goes out, the LED on the forehead keeps flashing,

and the equipment gets ready for measurement. If there is no operation for about 60 seconds, the

equipment will shut down automatically.

1.6 Placement

Place the thermometer between the eyebrows, at a distance within 5cm from the center of the

forehead;

In non-contact mode, the blue pilot light will point to the area you are aiming at. If the

eyebrow area is covered by hair, sweat or dirt, please clean it in advance to improve the accuracy

of the reading.

Keep the thermometer and forehead still when measuring, as movement will negatively affect

the temperature reading.

1.7 Mode switch

Toggle the slide switch to select the temperature measurement / calibration mode

1.8 Unit switch

In shutdown mode, press and hold the button "MEM”, it first displays “---- M”, which flashes

along with the symbol "M"; press and hold the memory button for more than 4 seconds, it displays

“---- C + human icon + sound icon"; press and hold the button for 6 seconds until °C or °F appears,

press the button "MEM” to achieve switch between °C and °F. It defaults to the current setting

after rebooting, while it will save previous settings in case of power failure.

Summary of Contents for LFR30B

Page 1: ...rehead thermometer and other components from the case with care and check them one by one against the packing list When the equipment is moved a different environment the difference in temperature or humidity may lead to condensation to it in which case no use is allowed before condensation disappears 1 2 Product photo 1 3 Display screen 1 Object mode symbol 5 Unit symbol 2 Temperature display val...

Page 2: ...time the backlight goes out the LED on the forehead keeps flashing and the equipment gets ready for measurement If there is no operation for about 60 seconds the equipment will shut down automatically 1 6 Placement Place the thermometer between the eyebrows at a distance within 5cm from the center of the forehead In non contact mode the blue pilot light will point to the area you are aiming at If ...

Page 3: ...rement to turn on the forehead thermometer align the thermometer sensor to the position between eyebrows within 5cm from the forehead press the button ON OFF measurement at this time the distance focusing light comes on quickly adjust the appropriate distance preferably when the focusing light spot is within the pea grain size when a bi is heard about 1S later no bi heard if the sound is disabled ...

Page 4: ...es Press the button Memory again to display the number of memory groups M icon and about 1 second later show the display memory value the symbol M flashes 1 13 Memory deletion In shutdown mode press and hold the button Memory the sound setting appears 4 seconds later at first and keep holding to clear the memory value 8 seconds later at this time it shows CLr M which flashes with the sound of Bi B...

Page 5: ...ized fee paying maintenance services 1 16 Measured temperature too high Error message on screen Meaning of show Possible causes and solutions Measured body temperature is too high above 43 0 C The result exceeds the measurement range limit Measured object temperature is too high above 100 C The result exceeds the measurement range limit 1 17 Measured temperature too low Error message on screen Mea...

Page 6: ...isinfect the product only when necessary disinfection is recommended to carry out by wiping the product surface twice with disinfectant Do not soak or use gas for disinfection Use of medical alcohol is recommended If necessary clean the infrared forehead thermometer first when repeated measurement of the body temperature is required Clean the thermometer shell and sensor with alcohol swab or cotto...

Page 7: ...easons as follows Repair service beyond warranty will be charged accordingly 1 Failure caused by unauthorized disassembly or refitting of the product 2 Failure caused by dropping during use or handling 3 Failure due to lack of proper maintenance 4 Failure to operate in accordance with the correct instructions in the operating manual Failure caused by failing to operate correctly as instructed by t...