Status LED 

Indicator 

No power, No connection, 
controller failure 

Off – relay open 

Control Operation 

 

< Low Set Point 

Orange  (constant) – relay closed  

Within Control Band 

Green    (constant) 

> High Set Point 

Red       (constant) – relay open 

Control Status 

 

Warm Up (30 minutes) 

Green  (Slow Flash -2 sec period) 
- relay open 

Wrong Switch Setting (eg.  
low rotary > high rotary) 

Orange (Fast Flash -0.5 sec period) – 
relay open 

Sensor Failure 

Red  (Fast Flash  0.5 sec period) 
– relay open 

Switch Operation 

 

Normal operation 

Green   (constant) 

Low Switch 

Yellow  (constant) 

High Switch 

Red      (constant) 

Sensor Head Failure 

The sensor failure LED signal (red fast flash) will activate due to: 

Sensor Life Exceeded –

this indicates that the sensor has 

operated beyond its design life and should be replaced.

 

Sensor Failure-

 

the sensor head is no longer operating 

within the on-board diagnostic settings and should be 
replaced.

 

16  

Summary of Contents for aeroQUAL

Page 1: ...User Guide Series 100 Ozone Controller Lenntech info lenntech com Tel 31 152 610 900 www lenntech com Fax 31 152 616 289 ...

Page 2: ...roller Components 10 Setting Up the Ozone Controller 11 Assembly 11 Power Requirements 11 External Control Wiring 12 Operating Instructions 13 Operating in Low or High Ozone Environments 13 Operating as a Controller 13 Setting the Control Set Points 14 Operating as a Switch 15 LED Status Indicators 16 Sensor Head Failure 16 Specifications Series 100 17 Drawings 18 ...

Page 3: ...products described in this document without prior notice Under no circumstances shall Aeroqual be responsible for any loss of data or income or any special incidental consequential or indirect damages howsoever caused The contents of this document are provided as is Except as required by applicable law no warranties of any kind either express or implied including but not limited to the implied war...

Page 4: ...ipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equi...

Page 5: ...table Products Warranty Period Series 100 Controller One year from the date of purchase High Ozone Concentration Head Six months from the date of purchase Low Ozone Concentration Head Six months from the date of purchase All accessories for the product are covered by a warranty for a period as follows Accessories Warranty Period Other Accessories One year from the date of purchase This warranty is...

Page 6: ...ich have been caused by use of non Aeroqual original accessories 6 This warranty becomes void if a non Aeroqual approved AC DC adaptor or battery is used 7 Tampering with any part of the product will void the warranty 8 Damage to the sensors can occur through exposure to certain sensor poisons such as silicones tetraethyl lead paints and adhesives Use of Aeroqual sensors in these environments cont...

Page 7: ...user guide for detailed safety instructions Do not connect incompatible products HAZARDOUS ENVIRONMENTS Do not use the Ozone Controller in or near volatile fuel or chemicals DO NOT EXPOSE SENSORS TO HIGH LEVELS OF OZONE This can adversely affect the calibration of Aeroqual sensor heads HEALTH AND SAFETY IN THE WORKPLACE The Aeroqual 100 Series Controllers are used to monitor Ozone concentration Ae...

Page 8: ... is designed for use at temperatures between 5 C and 50 C 23 F and 120 F Sudden changes in temperature will cause condensation that may damage the electronic componentry Do not attempt to open Non expert handling of the device may cause damage Do not drop knock or shake as this could lead to internal damage Do not use harsh chemicals cleaning solvents or strong detergents for cleaning Wipe with a ...

Page 9: ...rmanent Controller Placement The Aeroqual Ozone Controller has been designed to measure the ambient concentration of ozone The controller must not be placed directly in an ozone stream For indoor local area monitoring attach the controller to an inert fixture such as a metal post or rack with the inlet facing generally towards the centre of the room For leak detection mount the unit near the ozone...

Page 10: ...ed so that it does not normally require re calibration for the life of the sensor head The Aeroqual Ozone Controller comes with built in diagnostics which will alert the user if the sensor is not operating correctly Environmental conditions such as dust high humidity vibration chemicals and heat or cold as well as high concentrations of ozone may degrade the sensor performance and shorten the sens...

Page 11: ...o replace the controller 2 The High Concentration Ozone Sensor Head is designed to measure ozone concentrations from 0 00 to 5 00 ppm and is accurate to within 10 between 0 50 ppm and 5 00 ppm The T90 response time i e the time to reach 90 of the actual ozone concentration of the high concentration sensor head is 30 seconds Exposure of this sensor head to ozone concentrations higher than 5 00 ppm ...

Page 12: ...uency is dependant upon the application conditions This has to be carried out by fitting the sensor head onto a Series 300 or Series 500 monitor see Series 300 500 User Guide This procedure must be carried out in a low ozone environment 0 00ppm to 0 01ppm after which the unit can be returned to the S100 base Ozone Controller Components Series 100 Ozone Controller The following components are suppl...

Page 13: ...s 100 controller is designed to be powered by a 12VDC 400mA power supply The permissible voltage range is Max Voltage 15 VDC Min Voltage 11 VDC Ensure that the power supply to the unit is sized to account for the voltage drop across the Cat 5 cable so that it delivers 12VDC to the unit in line with the above voltage range Typical resistance of Cat 5 cable is 9 38Ohms per 100m at 20ºC The unit does...

Page 14: ...45 connector viewed from the front are numbered from 1 to 8 with pin number 1 being on the right The pin designations are as follows 1 GND 2 Relay switching signal 3 12V 4 Not used 5 Not used 6 GND 7 Signal showing power connected to the S100 8 12V 12 ...

Page 15: ...points dead band are set by adjusting the two rotary switches as shown in the diagram on the following page Both control set points must be operating i e set to a switch position between 1 and F for the unit to operate as a controller If either control set point is set in switch position 0 the controller will operate as a switch see section Operating as a Switch on page 15 The ozone concentration ...

Page 16: ...ange LED and turning the external trigger signal off when the measured value is above the high set point red LED While the ozone concentration is rising from below the low set point to the upper set point the external trigger signal remains on While the ozone concentration is falling from above the upper set point to the lower set point the external trigger signal remains off The LED glows green w...

Page 17: ...perated as a switch by turning one of the rotary switches to the 0 position Low Switch If the high rotary switch is set at the 0 position the unit activates the external trigger when the measured value is below the low set point Adjusting the low concentration rotary switch position sets the low set point see table on page 14 High Switch If the low rotary switch is set at the 0 position the unit a...

Page 18: ...gh rotary Orange Fast Flash 0 5 sec period relay open Sensor Failure Red Fast Flash 0 5 sec period relay open Switch Operation Normal operation Green constant Low Switch Yellow constant High Switch Red constant Sensor Head Failure The sensor failure LED signal red fast flash will activate due to Sensor Life Exceeded this indicates that the sensor has operated beyond its design life and should be r...

Page 19: ...5 00 ppm T90 response rate standard Low concentration Ozone Head 80 seconds High concentration Ozone Head 30 seconds Operating temperature range 5 C to 50 C Relative humidity range 5 to 95 Removable replaceable sensor heads Yes High low control or set points 16 position rotary switch External signal for control switching 12 VDC 150 mA Power Requirements User supplied 12 VDC 400 mA 11VDC min 15VDC ...

Page 20: ...Drawings 18 Lenntech info lenntech com Tel 31 152 610 900 www lenntech com Fax 31 152 616 289 ...

Reviews: