Warm up
Prior to operation the controller must be warmed up to burn off
contaminants on the sensor. When the controller is first switched
on it will warm up for 30 minutes prior to measuring Ozone.
Further to the above warm up time it is recommended that the
controller is run for 24 hours prior to use if it has been switched
off for more than 7 days.
Reset Routine
*
High ozone head only
For optimal accuracy with the high concentration head, the unit
should be periodically aligned with its operating environment
(frequency is dependant upon the application conditions). This
has to be carried out by fitting the sensor head onto a Series 300
or Series 500 monitor (see Series 300/500 User Guide). This
procedure must be carried out in a low ozone environment
(0.00ppm to 0.01ppm) after which the unit can be returned to the
S100 base.
Ozone Controller Components
Series 100 Ozone Controller
The following components are supplied with the Series 100
Ozone Controller:
•
Series 100 controller base
•
Ozone sensor head
•
User guide
•
3 rubber pads
•
4 Velcro dots