Configuration
Power Analyzer NORMA 4000, NORMA 5000
29
EO1111G REV G
to this fundamental. The low-pass filter is not located in the signal
path so that the input signal is not in any way interfered with.
Value in column
Filter
is highlighted.
Press
Enter
.
Select a value or
OFF
, depending on the expected
fundamental, and confirm with
Enter
.
The entered value, or
OFF,
is shown in column
Filter
.
Save the settings with
SAVE
in the desired configuration.
Value in column
SyOut
is highlighted.
Press
Enter
.
To activate output, select
ON
.
– or –
To deactivate output, select
OFF
.
Confirm with
Enter
.
The entered value is shown in column
SyOut
.
Save the settings with
SAVE
in the desired configuration.
Note
The synchronisation output is connected at the Sync-BNC plug on
the backside of the unit. The output signal is a TTL pulse with 5Volt.
Important
The BNC can be used either as input or output. As soon as the
BNC plug is switched to input (
EXT
sync source or
OFF
selected)
the sync output menu is automatically switched to
OFF
(disabled).
7.5 Adjust date and time
Note
Normally, date and time must be set only once, as they do not
change with different configurations.
Move the cursor to menu item
Clock Setup
and press
Enter
.
The
Clock Setup
menu is displayed. The value in column
Year
is highlighted.
Press
Enter,
select a year and confirm with
Enter
.
The selected year is displayed.
Move the cursor to the next field and repeat the above step
until the correct date and time are shown.
Menu field
Clock Setup
shows the time in hours, minutes and
seconds.
Configure signal
output
Adjust date and
time