8.7 Changing of all data settings with personalized CODE
With this function (FM, UM-Limit, Limit, beeper, ratio, R*, F*) limit- and set values can
be programmed which keeps them memorized even if the instrument is switched
ON/OFF. This feature enables the operator to create an instrument set-up with
customer defined settings according to the specific need.
Settings can only be made in the respective functions
(positon of central rotatry switch)
Function
Parameter
Setting range
Standard presetting
FM (AFC/94/105/111/128)Hz
AFC
RE 3pole
UM
48 V/20 V
48V
RK
0.000
Ω
... 29.99
Ω
0.000
Ω
and
LIMIT
On / Off
Off
RE LIMIT
0.000
Ω
... 999 k
Ω
999
k
Ω
RE 4pole
(Warning sound)
On/Off Off
R*
On/Off
Off
FM (AFC/94/105/111/128)Hz
AFC
RE 3pole
UM
48 V/20 V
48V
RK
0.000
Ω
... 29.99
Ω
0.000
Ω
and
I ( ratio )
80 ... 1200
1000
RE 4pole
LIMIT On/Off
Off
RE LIMIT
0.000
Ω
... 999 k
Ω
999
k
Ω
(Warning sound)
On/Off Off
R*
On/Off
Off
FM (AFC/94/105/111/128)Hz
AFC
RK
0.000
Ω
... 29.99
Ω
0.000
Ω
R~
LIMIT On/Off
Off
R ~ LIMIT
0.000
Ω
... 999 k
Ω
999
k
Ω
(Warning sound)
On/Off Off
R
...
2pole
RK
0.000
Ω
... 29.99
Ω
0.000
Ω
and
LIMIT On/Off
Off
4pole
R LIMIT
0.000
Ω
... 9,99 k
Ω
9.99
k
Ω
(Warning sound)
On/Off Off