S Series User Manual
Extend! 56
Installation of the Coded Zoom in LAS X (Continued)
4. Repeat the procedure to calibrate the knob
also for upper magnification (manually
set the zoom to highest magnification on
the microscope, then choose “Go to upper
limit”).
5. Repeat the procedure again to calibrate
the knob for 1.0× magnification (manually
set the zoom to 1.0× magnification on the
microscope, then choose “GO to nominal
1.0×”).
The coded zoom knob is now installed
and calibrated for your S9 microscope
and you can close the hardware configurator
tool.
Checking the functionality of the coded
zoom knob in LAS X
1. Open the LAS X software to check if the
coded zoom now operates correctly.
2. Navigate to the Magnification Menu.
3. Go to the microscope and turn the zoom
knob. The changing of the zoom should
now be reflected in the Magnification
Menu, and the zoom bar should move
accordingly.
All functionalities in the LAS X software
that correlate with magnification, e.g.
the scale bar or the measurement tool, use this
automatically detected magnification value
from the microscope for information.
Summary of Contents for S APO
Page 1: ...Leica S Series User Manual...
Page 10: ...S Series User Manual Leica S Series 10 Leica S Series...
Page 16: ...S Series User Manual Assembly 16 Assembly...
Page 26: ...S Series User Manual Quick Start Guide 26 Quick Start Guide...
Page 45: ...S Series User Manual Get Set 45 Get Set...
Page 52: ...S Series User Manual Extend 52 Extend...
Page 70: ...S Series User Manual Photography Video 70 Photography Video...
Page 73: ...S Series User Manual Dimensional Drawings in mm 73 Dimensional Drawings in mm...
Page 78: ...S Series User Manual Technical Data 78 Technical Data...