27
Leica RM2125
•
To mount the required insert (
62
,
63
,
64
) turn
the clamping ring (
66
) counterclockwise and
remove it.
•
Place the required insert into the clamping
ring (
66
) and screw the clamping ring onto
the thread (
69
) (rotate clockwise).
•
Insert specimen (
70
) and secure rotating the
clamping ring (
66
) clockwise.
•
To orient the specimen insert pin (
67
) into
hole (
68
) and turn counterclockwise to re-
lease the clamping mechanism. The speci-
men can now be oriented as desired.
•
Turn pin (
67
) clockwise to secure the speci-
men.
5.8.6 Holder for round specimens
The holder for round specimens accomodates cylindrical samples.
Inserts for cylindrical specimens of 6, 15 and 25 mm diameter are available.