Preparation before surgery
30
PROVIDO Ref. 10732418 / Version 03
PROVIDO Ref. 10732418 / Version 03 31
PROVIDO as far as the articulated arm.
X
Clamp the protective glass (optional) onto the objective.
X
Do not attach the sterile drape too tightly with the provided
ribbons. It must still be easy to move the instrument.
X
Check the ease of movement of the instrument.
Follow the instructions provided by the manufacturer of the
sterile drape.
Always use the drape with a protective glass.
Do not drape further than to position (1).
CAUTION
Risk of overheating of the system.
Covering the air inlet (2) can result in a controlled shutdown
of the system due to overheating.
X
Make sure the air inlet (2) is always kept free.
2
7.8.3 Attaching the protective glass to the objective
X
Place the sterilized protective glass on the optics carrier so that
the markings on the PROVIDO (1) and on the protective glass (2)
are aligned.
a
b
1
2
X
Insert the protective glass upwards into the bayonet mount in
direction (a).
X
Turn the protective glass in direction (b) until it engages.
7.9 Function check
Refer to the checklist before operation on page 72.