Emspira 3 User Manual Troubleshooting 51
Troubleshooting
Stand-alone operation
If Emspira 3 is configured for 4K resolution and
attached to a Full-HD monitor with only 1,080 p
resolution, Emspira 3 should detect this and
switch automatically to the necessary resolu-
tion.
However, if no image appears and the screen
reports an error message like "out of range" or
something similar, you can reset the resolution
to 1,080 p.
How to reset Emspira 3?
1. Power Emspira 3 off.
2. Unplug the USB mouse.
3. Restart Emspira 3.
Emspira 3 will restart in Full-HD 1,080 p
mode and a live image will appear.
4. Plug in the USB mouse again and configure
the correct live image resolution.