6. Assembly
6.12 Installation and Replacement
of Incident Light Lamps
Caution!
Light sources pose a potential irradiation risk
(glare, UV-radiation, IR-radiation). Therefore,
lamps have to be operated in closed hous-
ings and only after being mounted.
Ensure that the lamp housing has been dis-
connected from the power supply. Unplug the
power plug and the power supply during as-
sembly.
During assembly work on xenon burners,
always wear the supplied protective gloves
and face protection (Fig. 58) (risk of explo-
sion).
Never touch the glass parts of the burner
with bare hands.
Never look directly into the beam path (blind-
ing hazard).
Lamp Housing 106 z
This lamp housing is suitable for use with a 12 V
100 W halogen lamp or a variety of gas discharge
lamps.
Caution!
Make sure to follow the instructions and
safety notes of the lamp supplier.
Before changing lamps allow at least 30 min-
utes for cooling down!
Caution!
The lamp may still be hot.
Fig. 58
Protective gloves and mask
Fig. 59
Lamp housing 106 z L with Hg 100 W lamp
1
Collector focusing
2
Vertical lamp adjustment
3
Horizontal lamp adjustment
4
Hg lamp mount
5
Refl ector adjustment (not visible)
3
1
4
5
2
Summary of Contents for DMI Series
Page 1: ...Leica DMI3000B DMI4000B DMI6000B Instructions Bedienungsanleitung Mode d emploi...
Page 3: ...Leica DMI3000B DMI4000B DMI6000B Instructions...
Page 120: ......
Page 121: ...Leica DMI3000B DMI4000B DMI6000B Bedienungsanleitung...
Page 238: ......
Page 239: ...Leica DMI3000B DMI4000B DMI6000B Mode d emploi...