Messages and Warnings
12
Leica DI C800 / Ref. 10 743 250 / Version 00
Troubleshooting the mirrored data
Fault: No image superimposition.
Possible cause 1
: The data source does not emit an image.
The plug connection for the XGA video cable is defective or the XGA video cable is
connected incorrectly.
Fault remedy:
➩
Check whether the "DI C800" quick access button is present on the user interface.
➩
Check whether the video signal is correct.
➩
Check whether the data source is sending an image.
➩
Check the plug connection for the XGA video cable between the Leica DI C800
and the external data source.
➩
If the XGA video cable is connected correctly to the external data source, contact
your Leica representative.
Cause 2:
The Leica DI C800 does not have power.
Fault remedy:
➩
Check the plug connections for the power supply cable between the
Leica DI C800 and the optics carrier.
➩
If the power supply cable is connected correctly to the optics carrier, contact your
Leica representative.
Cause 3:
The Leica DI C800 is not enabled.
Fault remedy:
➩
Check whether the "DI C800" quick access button is present on the user interface.
➩
If the "DI C800" quick access button is present, contact your Leica representative.
Fault: The superimposed image does not have the correct color.
Cause:
defective plug connections.
The plug connection for the XGA video cable is defective or the XGA video cable is
connected incorrectly.
Fault remedy:
➩
Check the plug connection for the XGA video cable between the Leica DI C800
and the external data source.
➩
If the XGA video cable is connected correctly to the external data source, contact
your Leica representative.
Fault: The superimposed image is shown in the wrong format.
Cause:
The connected workstation is supplying data/signals in the wrong format.
Fault remedy:
➩
Check the graphics card settings on the workstation.
Fault: A correlated contour is not overlaid over the object precisely.
Cause:
The data source is not calibrated to the Leica DI C800.
Fault remedy:
➩
Have the data source calibrated to the Leica DI C800.
Troubleshooting the display
Fault: The display shows a signal error
Cause:
The display switches on by itself and/or shows an incorrect video signal.
Fault remedy:
➩
Unplug the Leica DI C800 and then plug it back in.
➩
Switch the Leica DI C800 off and then back on using the handswitch/footswitch
or the touchscreen.
Troubleshooting the optical system
Fault: Prolonged operation causes headaches for the surgeon
Cause:
Optics out of alignment.
Fault remedy:
➩
Contact your Leica representative.
Messages and Warnings