Check & Adjust
SPRINTER 100/100M/200/200M
47
7 Check & Adjust
In this chapter
Topic
Page
Description
Both optical and electronic collimation error can occur in the SPRINTER
100/100M/200/200M. Electronic staff readings are automatically corrected with the colli-
mation error stored in the SPRINTER 100/100M/200/200M while the optical collimation
errors need to be eliminated by adjusting the reticle. The instrument offers A x Bx field
procedure for the electronic measurements (A and B are the staffs positions and x is the
instrument position).
7.1
Procedures
Set up the SPRINTER 100/100M/200/200M in the center of the two staffs A and B. The
distance between the two staffs is approximately 30m.