14
Message Codes
de
GB
fr
it
es
nl
ja
*
Maximum deviation occurs under unfavorable condi-
tions, such as in bright sunlight or when measuring to
inadequately reflecting surfaces. For distances over
30m-without using a target plate- themaximum devia-
tion may increase by ±0.1mm/m to maximum ±
10mm. For more information, refer to
www.disto.com
.
** As of 100m the measuring units are displayed in cm.
*** Reduced when using the BLUETOOTH
®
function
Message Codes
Error message
appears beside the message number on the
display.
Please call your local dealership and specify the
message code. Should the message continue to
appear after several starts, then the instrument
is defective.
Care
Look after the optical surfaces with the same care that you
would apply to spectacles, cameras and field glasses.
The most current information can be found on our
homepage:
www.disto.com
We publish the addresses of specialized software
mannufacturers . However, we are not responsible for the
correctness of their specifications or for the performance
of their software.
For software developers, the Leica DISTO™ plus Online-
Interface description is available.
All illustrations, descriptions and technical specifica-
tions are subject to change without prior notice.
Message
Code
Cause
Remedy
203
Wrong sequence with
Pythagoras meas.
Carry out measurement in
correct sequence
204
Calculation error
Repeat procedure
240
Transmission error
Repeat procedure
252
Temperature too high,
above 50°C (measuring)
Cool down instrument
253
Temperature too low,
below -10°C (measuring)
Warm up instrument
255
Receiver signal too weak,
measurement time too
long, distance <200 mm
Use target plate
measurement time >10
sec.
256
Received signal too
powerful
Use target plate (correct
side)
257
Wrong measurement,
ambient brightness too
high
Use target plate
260
Laser beam interrupted
Repeat measurement
All other messages
Call service "System"
Message
Code
Cause
Remedy
Summary of Contents for DISTO plus