8
Instructions for Use V 1.4 RevC – 09/2013
3.
Instrument components and specifications
3.2 Technical data
General
Approvals:
The instrument-specific marks are located on the
name plate.
Operating temperature range:
10 °C to +40 °C
Relative humidity:
max. 80% non-condensing
Operating temperature range during storage: + 5 °C to +55 °C
Storage humidity:
< 80 %
Microtome
Section thickness range:
0.5 - 60.0 µm
Section thickness settings:
from 0.5 - 5.0 µm in 0.5 µm increments
from 5.0 - 10.0 µm in 1.0 µm increments
from 10.0 - 20.0 µm in 2.0 µm increments
from 20.0 - 60.0 µm in 5.0 µm increments
Automatic specimen feed:
from 0 to 30 µm
Total specimen stroke:
approx. 50 mm
Clearance angle adjustment:
–3° to 10°
Maximum specimen size:
50 x 60 x 40 mm
Specimen orientation
in cutting direction:
± 8°
perpendicular to cutting direction:
± 8°
Declination:
0° - 45° in cutting direction
Dimensions and weight
Width (with coarse driving wheel
and Ergo handle):
390 mm
Width (base plate):
256 mm
Depth:
430 mm
Height (total):
343 mm (with blade holder)
Working height (knife blade):
255 mm (measured from the table)
Weight (without accessories):
Approx. 20 kg