33
Leica CM3600 XP
Software
6
6.4.2 Time
Fig. 18
Displays real time and date.
•
See
p. 38 – 6.5.1 Parameter settings)
on how to configure these
settings.
6.4.3 Automatic defrost
11
Fig. 19
Displays start time of 1st or 2nd automatic defrost cycle.
•
See
p. 38 – 6.5.1 Parameter settings)
on how to configure these
settings.
Manual defrost
button
•
Clicking on this button
immediately starts a defrost
cycle.
•
During the defrost cycle, the button blinks yellow.
•
Clicking on the button once more aborts the defrost cycle.
6.4.4 Automatic dehydration
12
Fig. 20
Displays start time, start date and duration of the automatic dehydration.
•
See
p. 38 – 6.5.1 Parameter settings)
on how to configure these
settings.
Manual dehydration
button
•
Clicking on this button
starts the dehydration process.
•
Clicking on this button once more, stops dehydration.
•
When activated, the button blinks yellow.
•
If a reduction in temperature is to be carried out during automatic
dehydration, the required settings have to be made in the
Parameter
setting
window, see
p. 38 – 6.5.1 Parameter settings)
6.4.5 Knife movement
Displays section and retraction thickness. The
Actual position
displays the current
position of the knife holder. The
Set position
displays the set position of the knife holder.
•
The red
Retraction
display
only appears during the return stroke of the sledge.
•
To modify any of the settings, click on the corresponding icon. Entries are completed by pressing the
Enter
button.
Note
•
The knife movement buttons can only be used when the
Manual
operating mode has been set, see
.
•
The entry limit for the actual and target position can either be a maximum of 100,000 or