Performance Verification
DA18000AC-OM-E Rev A
7-3
Figure 7-1. Timebase Menu
h.
Connect the ST-20 TDR coaxial cable from Channel 1 of the WE 9000 via the LPA–K
adapter to Channel 1 of the SDA 18000. Be sure to properly torque all connectors
(8 in.-lbs.).
i.
Set the SDA 18000 to Channel 1 and the timebase to 500 ps/div and 60 GS/s. Set the
scope to trigger on Channel 1, and edge triggering to positive slope.
j.
Set math trace F1 to the
Ptrace mean
of Channel 1 (Figure 7-2).