Specifications
DA18000-OM-E Rev A
8-1
8
Specifications
The specifications are valid for probes when the following conditions have been met:
• The probe has been operating for at least 30 minutes in an environment, which is
within the operating environmental specifications.
• The probe has been calibrated within the last 12 months. Calibration was performed
in a controlled environment of 25 °C ±5 °C.
NOMINAL CHARACTERISTICS
Nominal characteristics describe parameters and attributes that are guaranteed by design,
but do not have associated tolerances.
General
Input Configuration
True differential, + and – Inputs
Input Connectors
2.92 mm
The 2.92 mm input leads will connect directly to
female 2.92 mm connectors on the system under
test.
Maximum Non-Destruct
Input Voltage, continuous
10 V
peak
Amplifier Gain
~ 1.2
Oscilloscope display factor is calibrated to actual amplifier gain
Input Impedance
100
Ω
, true differential
Max Input: Differential Mode
Without
÷
2 attenuators
±400 mV, 800 mV
p-p
With
÷
2 attenuators
±800 mV, 1.6 V
p-p
Max Input: Common Mode
Without
÷
2 attenuators
10 mV
p-p
With
÷
2 attenuators
7 V
rms
WARRANTED CHARACTERISTICS
Warranted characteristics describe parameters that have guaranteed performance. Unless
otherwise noted, tests are provided in Section 8, Performance Verification, for all warranted
specifications.
Rise Time (Amplifier only)
1
19 ps
(20-80%)
Voltage Gain Accuracy
> 2%
1. Measured with an SDA 18000 oscilloscope. Rise time of oscilloscope and source sub-
tracted (rms).
Summary of Contents for DA18000
Page 1: ...LeCroy DA18000 Differential Amplifier Instruction Manual December 2006 ...
Page 4: ...iv DA18000 OM E Rev A BLANK PAGE ...
Page 10: ...DA18000 Differential Amplifier 2 2 DA18000 OM E Rev A BLANK PAGE ...
Page 12: ...DA18000 Differential Amplifier 3 2 DA18000 OM E Rev A BLANK PAGE ...
Page 22: ...DA18000 Differential Amplifier 6 4 DA18000 OM E Rev A BLANK PAGE ...
Page 34: ...DA18000 Differential Amplifier A 2 DA18000 OM E Rev A BLANK PAGE ...