Warning Alarms

This fridge-freezer has been designed to provide both an audible and visual alarm should any
of the following situations occur:

Door is left open for a set period

Internal air temperature goes outside the set parameters, either too high or too low

Internal ‘load’ temperature goes outside the set parameters, either too high or too low

Should one of these alarms occur this can be muted by pressing

on the controller.

The ‘load’ temperature is a simulated temperature which is intended to provide an indication
of the temperature of the fridge-freezer’s contents (e.g. a 5ml vaccine). The temperature of a
fridge-freezer’s contents will change more slowly (up or down) than the air temperature
within that fridge-freezer should there be a fault or the door is left open.

Remote alarm contacts

All Lec Medical laboratory fridge-freezers include remote alarm contacts and these terminal
contacts enable the user to connect to a remote control station; these contacts can be found on
the rear of the cabinet.

WARNING

: These contacts must only be accessed by a competent

electrician.

In order to use these contacts the following must be carried out:

1. The fridge-freezer MUST be disconnected from the power supply before removing

the terminal cover.

2. Unscrew the cover (figure 1) on the right hand side of the rear of the cabinet to expose

the connectors (figure 2).

3. Use the right hand outer positions that correspond to the left hand cable positions for

your connections using 0.75mm cable stripped to 0.6mm.

4. Depress the larger slot above the cable entrance on the right hand side with a small

electrical screwdriver. This allows the cable to be inserted into the slot directly below.

In the event of a door open alarm, high or low temperature alarm or power failure a remote
relay will switch within the controller to make an electrical circuit.

Please note that the contacts supplied are voltage free. A voltage of between 12V DC and
230V AC can be connected at the contacts. The maximum load must not exceed 2A. The
minimum power rating is 500mA / 12V AC.

Figure 1

Figure 2

Summary of Contents for LSC119

Page 1: ...1 Laboratory fridge freezer range Applicable models LSC119 LSC263 LSC324 USER MANUAL...

Page 2: ...verview 6 Viewing and resetting minimum and maximum temperatures 7 Viewing the freezer temperature 8 Changing the factory set point 8 Changing the temperature scale 9 Changing the remote alarm contact...

Page 3: ...placed in a dry well ventilated site away from heat sources The appliance must be placed on a level surface If necessary adjust the feet on the product so the appliance remains level during operation...

Page 4: ...fridge freezer Also make sure there is good ventilation in the room If you are in doubt please contact your supplier WARNING Keep ventilation openings in the appliance enclosure or in the built in str...

Page 5: ...sket 6 Adjustable foot 2 at front of appliance 7 Door furniture There are also 2 roller castors at the rear of the appliance to ease product movement not on LSC263 Additional or replacement parts are...

Page 6: ...6 Controller overview and operation 1 Menu 2 Lock keypad 3 Maximum temperature Up 4 Power 5 Mute 6 Minimum temperature Down 7 Set Enter 1 2 3 4 5 6 7...

Page 7: ...7 Viewing resetting minimum and maximum temperatures applies to fridge only...

Page 8: ...e circumstances such as high low room temperatures it may be necessary to adjust this slightly Function Press Hold button s down for Example screen To change the set point 3 seconds followed by Once f...

Page 9: ...lowed by Two times followed by 1 second current scale followed by Once followed by 1 second Changing the remote alarm contacts Function Press Hold button s down for Example screen To change the alarm...

Page 10: ...emote control station these contacts can be found on the rear of the cabinet WARNING These contacts must only be accessed by a competent electrician In order to use these contacts the following must b...

Page 11: ...11 In a normal condition i e no alarm the relay OPENS the contact for the remote alarm facility In an alarm condition the relay CLOSES the contact for the remote alarm facility...

Page 12: ...may be required Before manually defrosting the cabinet make sure the power lead is disconnected from the mains Any loose frost found on the evaporator can be removed carefully using a wooden or plast...

Page 13: ...rasives or acids are not suitable for the cleaning When all cleaning has been completed and accessories replaced reconnect the plug with dry hands Reversing the door In order to reverse the door pleas...

Page 14: ...site side of the door 14 Screw the handle into position and add the caps to cover the screws Storage If the fridge freezer is to remain unused for a period of time e g during refurbishment work switch...

Page 15: ...APPLIANCE MUST BE EARTHED The flexible cord mains lead fitted to this appliance has three cores for use with a 3 pin 13 amp plug If a BS 1363 13 amp fused plug is used it should be fitted with a 13 am...

Page 16: ...tains insulation gases and refrigerant and must be disposed of in the appropriate manner Please contact the manufacturer or local authority for advice on how best to dispose of this product A relevant...

Page 17: ...heck that the appliance is operating at an ambient temperature that meets the product specification on page 4 Air circulation at the back of the appliance is hampered 1 Move appliance further away fro...

Page 18: ...48 Marketing Enquiries email marketing gdpa co uk tel 0844 248 4661 Spares and After Sales Service email info gdha com tel 0844 815 3755 fax 0844 248 4123 Address Lec Medical Stoney Lane Prescot Merse...

Reviews: