Lazurite ArthroFree Instructions For Use Manual Download Page 33

33

The ArthroFree System

Warnings, Cautions, and Notes

Please ensure that all staff and medical professionals using this equipment read and understand this manual 
thoroughly. Failure to do so may result in incorrect operation and lead to patient harm. 
Warnings, cautions, and notes are listed here in the order they appear above. 

Warnings

To avoid potentially serious injury to the patient, clinician, or user. and to reduce the risk of damage to the 
ArthroFree System, please take note of the WARNINGS identified throughout this document as summarized in 
Table 4 below. The warranty is void if any warning is disregarded.

TABLE 5: Warnings 

Section

Warning

General 

Warnings

Before installing and using the ArthroFree System, read this manual thoroughly, especially 

the Warnings, Cautions, and Notes.

Product 

Description 

and Intended 

Use

The light source contains a Class 1 laser product. The light source is designed to prevent 

danger from the laser enclosed within it. Do not disassemble the light source. Use of the 

enclosed laser without the protective design is potentially hazardous. Do not look directly 

into the light source while disconnected or through a scope.

Do not short-circuit Batteries by bridging terminals with conductive materials, or open, 

crush, or incinerate Batteries. Doing so may lead to shock, burn, damage to Battery, or 

user injury. Failure to follow these instructions will void the warranty.

Only charge the Battery with the Battery Charger provided in the ArthroFree System. 

Failure to follow these instructions may result in damage to the Battery; damage to the 

Battery Charger; a fire or explosion; and/or injury to the patient, clinician, or user. Failure 

to follow these instructions will void the warranty.

System Setup

Carefully unpack all ArthroFree Components and verify that no damage occurred during 

shipment. Verify that all parts are present and match those on the included packing list.

If the packaging of the ArthroFree System (1) is damaged in any way prior to receipt 

by the customer, (2) has been opened prior to receipt by the customer, or (3) has been 

exposed to environmental conditions outside of those specified in Figure 9, contact 

Lazurite technical support before using.

The ArthroFree System is not shipped sterile. The Camera Head and Batteries must be 

properly sterilized prior to use in surgery (see Cleaning and Sterilization).

Do not position the Receiver within the sterile field.

Operating 

Instructions

Before using the ArthroFree System in a surgical procedure, test all Components to 

ensure proper function. To test the Camera Head, power it on by inserting a Battery and 

then point it at an object and ensure that a video image of that object appears on all 

connected surgical displays.

The ArthroFree user must be a qualified healthcare provider and must be familiar with 

the instructions for use of the equipment and any applicable surgical techniques.

Do not position the Battery Charger within the sterile field.

Do not power on more than one ArthroFree Camera Head within range of a single Receiver.

Summary of Contents for ArthroFree

Page 1: ...WARNING Before installing and using the ArthroFree System read this manual thoroughly especially the Warnings Cautions and Notes Instructions for Use Cleaning and Sterilization...

Page 2: ...The ArthroFree System Contact Information For technical assistance or sales support contact Lazurite at 1 833 214 2324 Lazurite 4760 Richmond Road Suite 400 Cleveland OH 44128 United States of Americ...

Page 3: ...Charger 10 ArthroFree Receiver 11 ArthroFree Sterilization Tray 11 System Compatibility 13 System Setup 14 Operating Instructions 18 Wireless Visualization 23 Cleaning and Sterilization 24 Camera Head...

Page 4: ...oubleshooting Troubleshooting 37 38 FIGURES Section Page Figure 1 ArthroFree Camera Head ArthroFree Battery Pack Product Description and Intended Use 8 Figure 2 Control Buttons Product Description and...

Page 5: ...arnings may result in injury to the patient clinician or user CAUTION indicates risks to the equipment Failure to follow Cautions may result in product damage NOTE indicates special information to cla...

Page 6: ...Indicates the manufacturer s serial number for this specific device Fragile Handle with care Indicates a medical device that can be broken or damaged if not handled carefully Temperature limit Indicat...

Page 7: ...ations The ArthroFree System is indicated for use in arthroscopy general laparoscopy nasopharyngoscopy ear endoscopy sinuscopy plastic surgery or wherever a laparoscope endoscope or arthroscope is ind...

Page 8: ...0001 101 10 ArthroFree Receiver AF0002 101 10 ArthroFree Battery AF0003 101 10 ArthroFree Battery Charger AF0004 101 10 ArthroFree Sterilization Tray AF0005 101 10 ArthroFree Wireless Camera Head The...

Page 9: ...existing C mount coupler and a scope The control buttons manage brightness digital zoom and white balance They enable still image and video capture when connected to a compatible pre existing capture...

Page 10: ...inserting it completely into one of the five charging bays on the Battery Charger While a Battery charges the corresponding OLED display indicates its charging progress and the number of charging cyc...

Page 11: ...or initiation of still photo and video capture on an existing data console Sterilization Tray When shipped as a kit or a set the Camera Head and three Batteries are packed in the Sterilization Tray Th...

Page 12: ...12 The ArthroFree System FIGURE 6 ArthroFree Sterilization Tray Light Cable Light Source Batteries 3 Wireless Camera Head Lid...

Page 13: ...patibility of the ArthroFree System with surgical tower components must be validated in clinical practice prior to first use To validate display a test image on the surgical display see Operating Inst...

Page 14: ...main surgical display or the surgical tower 1 EITHER place the Receiver on top of the pre existing surgical display as shown in Figure 7 by a First connecting the provided VESA mount bracket to the ba...

Page 15: ...ArthroFree Receiver 100mm VESA Mounting Width 200mm VESA Mounting Width Receiver Receiver Receiver Locking Screw Receiver Mounting Bracket VESA Mounting Bracket Receiver Receiver Mounting Bracket VESA...

Page 16: ...port on the pre existing patient data console After installation test for picture triggering and adjust port selection as necessary WARNING Do not position the Receiver within the sterile field CAUTI...

Page 17: ...ectrical outlet Position the Battery Charger so that the plug is easily accessible CAUTION When setting up the Battery Charger use only supplied cables and accessories Any other cables or accessories...

Page 18: ...all Components for signs of damage or wear Any signs of damage including cracking hazing chips nicks scratches and abrasion on the Camera Head or battery housing should be reported and immediately ass...

Page 19: ...light post of a pre existing endoscope Only pull from the main body of the light source Battery Insertion Removal and Replacement 1 To insert the Battery place it completely into the battery bay of th...

Page 20: ...NG The Battery can be stored in the Battery Charger for as long as required 1 To insert a Battery line up the Battery with an open charging bay and gently insert the Battery fully until the Battery st...

Page 21: ...record video of the wirelessly transmitted signal A short press of the camera button will activate remote trigger port 1 on the Receiver typically causing the capture of a still image of the displayed...

Page 22: ...tery charge 0 segments Indicates Battery is below 1 3 of the maximum charge level Replace with a fully charged Battery The icon flashes when the Battery reaches a critically low charge level Wireless...

Page 23: ...tical for successful use of the ArthroFree System NOTE The ArthroFree System is designed to provide a live video within ten feet of the Receiver with a clear line of sight between the Receiver and Cam...

Page 24: ...Do not place any Component of the ArthroFree System in a moist heat sterilizer autoclave or automated washing device Temperatures in these units may cause damage to sensitive electronics or the lithiu...

Page 25: ...at any point in their life or use cycle must be considered to be nonsterile and a potential source for disease transmission Any visibly soiled articles should be recleaned and sterilized prior to use...

Page 26: ...es into the Sterilization Tray in the locations visually indicated by the image in the base of the Sterilization Tray 3 Wrap the Sterilization Tray in a double layer of sterilization wrap 3 4 Use a pr...

Page 27: ...Batteries stored longer than 30 days should be recharged and reprocessed prior to use Repeated cleaning and sterilization may impact the expected service life of ArthroFree Components Inspect all Comp...

Page 28: ...umber of times a Component in the ArthroFree System is used cleaned or sterilized Actual service life is determined by factors such as handling wear sterilization methods chemical exposure and time be...

Page 29: ...dance with all applicable state and local regulations The Camera Head Receiver and Battery Charger contain electronic equipment Such Components must not be disposed of as unsorted municipal waste and...

Page 30: ...4 1V max charging voltage 100 240 VAC 50 60 Hz XX A Not applicable Device weight 371g 294g 83g 800g 1400g Classification Class II equipment Class II equipment Class II equipment Unclassed Class II eq...

Page 31: ...y is restricted to repair or replacement by Lazurite at its option of any product found to be defective during the warranty period Damage inflicted on a product by the user will result in additional c...

Page 32: ...s FCC ID 2AZYQ 01A This device complies with Part 15 of FCC Rules Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept an...

Page 33: ...Battery with the Battery Charger provided in the ArthroFree System Failure to follow these instructions may result in damage to the Battery damage to the Battery Charger a fire or explosion and or in...

Page 34: ...y point in their life or use cycle must be considered to be nonsterile and a potential source for disease transmission Any visibly soiled articles should be recleaned and sterilized prior to use with...

Page 35: ...e any ArthroFree Component in liquid Doing so may irreparably damage the Component Disconnect the Battery Charger and Receiver from the AC power source before cleaning Never clean the charging contact...

Page 36: ...ource is designed to connect to an ACMI snap in groove on the light post Any pre installed light post adapters corresponding to other attachment standards should be removed prior to connecting the lig...

Page 37: ...ersists contact Lazurite technical support Picture is too dark Increase brightness by quickly pressing and releasing the multifunction button on the Camera Head until the brightness icon is displayed...

Page 38: ...onnect before restarting surgery Battery fails to charge or is not detected Place Battery into different charger bay Replace Battery The Receiver indicator light is off Check that power is being deliv...

Page 39: ...hnologies See www lazurite co patents for more details Lazurite 4760 Richmond Road Suite 400 Cleveland OH 44128 United States of America 1 833 214 2324 1900D 20221 0 2022 Copyright Lazurite Lazurite A...