The ArthroFree System
2. Once the Receiver is in place, connect at least one of the provided DVI cables to one of the DVI-out
connections on the back of the Receiver and to a DVI input on the pre-existing patient data console.
If the Receiver is placed further than ten feet away from the pre-existing patient data console, use a
longer DVI cable.
3. Connect the provided power cable to the AC power port on the back of the Receiver and plug in male
end of power cord to a hospital-grade electrical outlet.
4. Connect the video trigger cable to remote trigger port #1 on the back of the Receiver and to the video
remote trigger port on the pre-existing patient data console. After installation, test for video triggering
and adjust port selection as necessary.
5. Connect the picture trigger cable to remote trigger port #2 on the back of the Receiver and to the
picture remote trigger port on the pre-existing patient data console. After installation, test for picture
triggering and adjust port selection as necessary.
WARNING: Do not position the Receiver within the sterile field.
CAUTION: The Receiver requires electricity via a hospital-grade electrical outlet. Position the
Receiver so that the plug is easily accessible.
CAUTION: When setting up the Receiver, use only supplied cables and accessories. Any other
cables or accessories may result in increased EMC emissions (See Technical Specifications:
Electromagnetic Compatibility).
Battery Charger Installation
1. Place the Battery Charger on a flat, secure, and stable surface in a well-ventilated and secure location.
2. Ensure that there are no foreign substances in the charger bays.
3. Connect the AC power cord to a hospital-grade outlet. Ensure that this outlet is not being used for life-
support equipment.
4. Each charger bay should now indicate "NO BATTERY" on its respective OLED display.
FIGURE 8: ArthroFree Receiver Connections
DVI-out (2)
AC Power
Remote Trigger Port #1: Long Press
Remote Trigger Port #2: Short Press