Specifications
OX-TRAN Model 2/22 Operator’s Manual
9-4
Revision R
MOCON, Inc.
Temperature Control Range
10 °C to 40 °C ± 0.2 °C
Time to Reach a Heated Temp (from ambient) Approximately 45 minutes to reach 99% of a 40 °C setting
Time to Reach a Cooled Temp (from ambient) Approximately 2 hours to reach 99% of a 10 °C setting
Humidity Control, Film Cell, Carrier Gas
0% to 90% RH ± 3 RH
Humidity Control, Film Cell, Test Gas
0% to 90% RH ± 3 RH
RH Film Cell-to-Cell Uniformity
± 3 RH
Time to Reach Humidity Setting
Approximately 3 hours to reach 99% of RH setting
Test Range, Film, 50 cm
2
0.0005 to 200 cc/(m
2
• day)
0.000032 to 12.9 cc/(100in
2
• day)
Test Range, Package
0.000003 to 1.0 cc/(pkg • day)
Resolution
0.0001 cc/(m
2
• day)
0.000006 cc/(100in
2
• day)
0.0000005 cc/(pkg • day)
Repeatability
± 0.0005 cc/(m
2
• day)
± 0.000032 cc/(100in
2
• day)
± 0.000003 cc/(pkg • day)
or 1% of value, whichever is greater
Table 9.8: Operational Capabilities - Ox-Tran 2/22 Model 10X