- 5 -

INSTRUMENT START-UP: 

 

Switch on the instrument and wait for about 5 minutes until all the components will reach a steady 
temperature. Verify that the displayed weight is positive and that it increases if one subjects the 
container to a force weight. In case a negative value is displayed check the load cells connections 
and their correct installation ( direction of charge ). Check also that the container is correctly 

installed ( tubes position, links, etc.). 
If the “JOL.4” message appears, select the desired operating modesee page  1. 

THEORETICAL INSTRUMENT CALIBRATION

:  

 

Switch off the instrument, then turn on by keeping pressed the MENU key, the C.0.S.c.”, message 

appears, press ENTER and the following will appear: 

“dECP”  , set the number of decimals (max. 3 decimals) using the

and 

 keys. Confirm 

with ENTER , the following will appear: 

“CELL”  , set 10% of the system’s full scale by using the 

 and 

 keys (for example with 3 

load cells of 100 kg ; full scale = 100 x 3 = 300.0 kg ; 10% of 300.0 kg = 30.0 kg). 
Confirm with ENTER , the following will appear: 

“nU-U” , set the load cell sensitivity expressed in mV/V (normally 2,000) using the

 and 

  

keys. Confirm with ENTER , the following will appear: 

“rISO” , set the system resolution by using the

 and 

  keys: x 1, 2, 3, 4, 5 . Confirm with 

ENTER , the following will appear: 

 “unit”, set the unit of measure using the

and 

 keys: 0 = kg;  1 = t. 

 Confirm with ENTER , the following will appear: 

“tArE”  , set the tare value, if you knows it, by using the

and 

 keys. Confirm with ENTER 

and C.0.S.c.” message appears again. 

 

Press MENU to exit. 

 
 
TARE ZERO-SETTING: 
Make sure that the container is empty and press the

 key, the "tArE " message will be displayed, 

keeping pressed 

 press the ENTER key at the same time and the tare will be set to zero. 

By pressing 

 , after 1 second, the tare value appears. 

 

It is also possible to set the tare to the zero by using the ZERO-SETTING EXTERNAL 

CONTACT (ONLY FOR “4SEt” OPERATING ); in this case the LED of DOSA key will start 
blinking. 

 

WARNING: TURNING OFF THE INSTRUMENT, THE DIVISIONS SET TO ZERO BY 
EXTERNAL CONTACT WILL BE LOST.  

 

 

 

Summary of Contents for JOLLY4

Page 1: ...user s manual version 3 3 JOLLY4 4 different operating modes selectable 2014 30 EU EMC 2014 35 EU LVD DATA TO BE FILLED OUT BY THE INSTALLER Page 1 ENGLISH ENGLISH ENGLISH ENGLISH ENGLISH ENGLISH ENG...

Page 2: ...ication label and or serial number of the product misuse transformation alteration repair of products not carried out by Laumas personnel Laumas provides a 1 year warranty from the date of the deliver...

Page 3: ...V INPUTS OUTPUTS 4 INSTRUMENT START UP 5 THEORETICAL INSTRUMENT CALIBRATION 5 TARE ZERO SETTING 5 INSTRUMENT CALIBRATION CHECK AND CORRECTING DISPLAYED WEIGHT 6 WEIGHT INDICATOR WITH FOUR ALARM THRESH...

Page 4: ...of 4 products Then press ENTER to restore the instruments the selected operating mode appears followed by the version s number If you press ENTER while JOL 4 0 is displayed the JOL 4 message will appe...

Page 5: ...0 ohm or max 8 load cells 700 ohm INTERNAL DIVISIONS 20000 DISPLAY RANGE 999 19999 MEASURING RANGE 4 mV 16 5 mV READING RESOLUTION x 1 x 2 x 5 CONVERSION RATE 10 readings sec LOAD CELLS SIGNAL READING...

Page 6: ...having the same colour perform the same operation for the shields By means of a shielded 4 wire cable with a section not lower than 0 5 mmq connect the terminal board to the JOLLY instrument The cabl...

Page 7: ...message appears press ENTER to close open the relay 1 by ensuring the changeover from to _ _ _ when the relative input changes from open to closed Press the 2 message appears press ENTER to close open...

Page 8: ...ells of 100 kg full scale 100 x 3 300 0 kg 10 of 300 0 kg 30 0 kg Confirm with ENTER the following will appear nU U set the load cell sensitivity expressed in mV V normally 2 000 using the and keys Co...

Page 9: ...the cell and the cell loading position If the difference is not significant but around 1 2 then proceed to CORRECTING DISPLAYED WEIGHT FROM KEYBOARD CORRECTING DISPLAYED WEIGHT FROM KEYBOARD When the...

Page 10: ...he weight for SET 1 max 9999 by using the and keys then confirm with ENTER the following will appear SEt 2 program the weight for SET 2 by using the and keys then confirm with ENTER the following will...

Page 11: ...PAUS time has elapsed the instrument closes the cycle end contact led 4 the gross weight will appears When the unloading phase starts when the minimum weight is reached and the safe time emptying time...

Page 12: ...ce the weight oscillations program a value included between 0 1 and max 9 9 seconds Confirm with ENTER to exit By pressing MENU you can exit in any moment 3CAr PROGRAM Loading batching of 3 products C...

Page 13: ...top of the extractor 0 manual fall or 1 automatic fall The instrument will calculate on itself the fall value by using the arrows keys then confirm with ENTER the following will appear VOL 1 program t...

Page 14: ...ontact led 2 when the SET2 value less the vol2 value is reached the SET2 contact will be opened after PAUS time has elapsed the instrument executes the autotare and closes the SET3 contact led 3 when...

Page 15: ...r VOL 2 value of fall 2 Confirm with ENTER the following will appear VOL 3 value of fall 3 Confirm with ENTER the following will appear VOL 4 value of fall 4 Confirm with ENTER the following will appe...

Page 16: ...product manuals are available online You can download the manuals in PDF format from www laumas com by consulting the Products section or the Download Area Registration is required Think about the en...

Reviews: