Personnel:
n
Operating personnel
1.
Change to the main menu.
2.
Select the menu item
Setup
Basic setup
Language
.
3.
Select any of the available languages.
4.
Confirm with the OK button.
6.13 Entering the offset of the internal actual temperature (calibration)
The factory calibration is overwritten during the adjust-
ment. This requires a reference thermometer with the
required level of accuracy. The factory calibration should
otherwise not be changed.
A temperature deviation can be corrected if inspection of the ther-
mostatic circulator with a calibrated reference thermometer shows
such a deviation.
The probe of the reference thermometer must be installed in the
outflow of the device in accordance with the specifications in the
calibration certificate.
Personnel:
n
Operating personnel
1.
Change to the main menu.
2.
Select the menu item
Setup
Calibration
Calibration
.
3.
Adjust the value accordingly. The value displayed on the ref-
erence thermometer must be entered.
4.
Confirm with the OK button.
6.14 Restoring factory calibration (internal temperature sensor)
Any offset specified for the internal temperature measurement can
be restored to the factory setting.
Fig. 49: Selecting menu language
Fig. 50: Specifying offset
Operation
Variocool
67