Command
Meaning
RMP_START
Starts programmer
RMP_PAUSE
Stops programmer
RMP_CONT
Starts programmer again after hold
RMP_STOP
Stops the program
RMP_RESET
Deletes program (all segments)
RMP_OUT_00_XXX.XX_XXX
XX_XXX.XX_X
Sets the programmer segment (temperature, time, tolerance and pump
level). A segment is added and allocated the appropriate values.
RMP_OUT_02_XXX
Number of program cycles: 0 = infinite / 1–250
Please note:
n
For “_” “.” (space character) is also permitted.
n
Response from the thermostat “OK” or “ERR_X” in the case of
an error. RS
‑
485
‑
interface e.g. “A015_OK” or with error
“A015_ERR_X”.
n
The command from the control centre must be terminated with
CR, CRLF or LFCR.
n
The response from the constant temperature equipment is
always terminated with CRLF.
n
After every command sent to the constant temperature equip-
ment, it is necessary to wait for a response before the next
command can be sent. This way, the allocation of queries and
responses is clear.
CR = Carriage Return (hex: 0D); LF = Line Feed (hex: 0A)
-XXXX.XX
-XXXX.X
-XXXX.
-XXXX
XXXX.XX
XXXX.X
XXXX.
XXXX
-XXX.XX
-XXX.X
-XXX.
-XXX
XXX.XX
XXX.X
XXX.
XXX
-XX.XX
-XX.X
-XX.
-XX
XX.XX
XX.X
XX.
XX
-X.XX
-X.X
-X.
-X
X.XX
X.X
X.
X
-.XX
-.X
.XX
.X
4.4.7 Error messages from the constant temperature equipment to the control
centre
This list describes the error messages.
The following information refers to the Ethernet
‑
as well
as the RS 232/485
‑
interfaces.
Permitted data formats
Before starting up
Variocool
36