LAUDA Ultra-Thermostats
UB 20(-D), UB 25, UB 50, UB 30, UB 40
UB 20 F, UB 20 J, UB 30 J, UB 40 J, UB 65 J
UB 20 JL, UB 30 JL, UB 40 JL
YATE0013 /10.04.02
- 10 -
2.2
Controller R 400-P
Ambient temperature
(°C)
5...40
Control range
(°C)
-65...305
Temperature setting/
Resolution
(°C)
membrane keypad with 16 keys, setpoint input with 0.01°C resolution
Temperature measurement (bath)
built-in digital thermometer with 0.01°C resolution; accuracy and stability of measurement
(electronics without probe) better than 0.05 %
±
0.05 K*).Temperature probe Pt 100 to
DIN IEC 751, Class B, can be calibrated additively at each measurement point.
External temperature measurement
2 separate temperature measurement circuits for external Pt 100 to DIN EN 60751 in 4-wire
circuit, accuracy and stability of measurement better than 0.05 %
±
0.05 K*),
can be calibrated additively at each measurement point. One probe produces the
measurement for external control.
Display
back-lit LCD matrix display, 2 lines with 16 signs each, 10 mm character height
Temperature control
modified PID controller with automatic structure selection. Control parameters through auto-
adaptation or manual input.With external control a cascade controller operates on the
measurement of one of the two external circuits (T
1
or T
2
).
Heating actuator
triac zero crossing with full wave PWM,
capacity max. 3 kW or 14 A
Triggering for
triac control of control valve for cooling water solenoid valve 230 V; 50/60 Hz, max. 0.2 A
controlled cooling
Safety devices
In the working temperature range adjustable overtemperature protection and adjustable low-
level protection.
Pump and heater are switched off on all poles.
Overtemperature measurement
Pt 100 in the thermostating part
Terminal strip for
overall fault, input fault, analogue signals, 2 x Pt 100 Ext
interface
Digital interface
RS 232 C
Bench area (W x D x H)
(mm)
190 x 200 x 180
Weight
(kg)
4
Supply
230 V; 50/60 Hz
Loading (max.)
3.2 kW
without accessories
current take-up (max.)
16 A
with accessories
Cat. No.
(only if controller is ordered separately)
R 400-P
LRK 013
*) see Item 4.3
Technical changes reserved.