Proline Kryomats
122
Technical data
10/2/2018
11
Technical data and diagrams
The figures have been determined according to DIN 12876.
Table 1
RP 3050 C
RP 3050 CW
RP 4050 C
RP 4050 CW
Operating temp.- ACC range
°C
-50
– 200
Ambient temp. range
°C
5
– 40
Relative humidity
maximum relative humidity 80 % for temperatures up to 31 °C,
decreasing linearly to 50 % relative humidity at 40 °C
Device distance to the
surroundings
cm
50
20
50
20
Temperature range for storage
°C
-20
– 44
the condenser must be completely emptied by a water-cooled device (
Setting resolution
°C
0.1 / 0.01 (Master); 0.01 (Command)
Display resolution
°C
Master: 0.01 Command: 0,1 / 0,01 / 0,001
Display accuracy
±0.2 °C can be calibrated additively (
Section 1.3 last Point)
Temperature stability
@ -10 °C with ethanol
K
±0.05
Safety equipment
Class
III, FL suitable for flammable and non-flammable liquids
Cooling
Air
Water
Air
Water
Ccooling water consumption:
temperature 15 °C, pressure 3
bar
L/h
700
700
Water-cooling connections
(DIN EN 10226-1)
inch
G ¾”
G ¾”
Heater power 400 V
kW
maximum 3.5
Heater power 208 V
kW
maximum 3.0
Heater power 200 V
kW
maximum 2.8
Cool
in
g p
o
wer
at
20
°C t
am
b
@
ba
th
tem
p
. (Pu
mp
Le
v
e
l
6)
with heat
transfer oil
200 °C
kW
5.0
6.0
5.0
6.0
with ethanol
20 °C
kW
5.0
6.0
5.0
6.0
0 °C
kW
3.0
3.5
3.0
3.5
-20 °C
kW
1.6
1.8
1.6
1.8
-30 °C
kW
1.0
1.1
1.0
1.1
-40 °C
kW
0.5
0.6
0.5
0.6
-50 °C
kW
0.25
0.25
0.25
0.25
Pump type
Pressure pump, 4 power levels (level 5 to 8)
Discharge pressure max.
bar
0.5 at pump power level 8
Flow rate max. (pressure)
L/min
19 at pump power level 8
Hose connections
Thread M16 x 1; olives 13 mm external diameter
Bath volume from
– to
L
23
– 31
32
– 44
Bath opening B x L
mm
350 x 200
350 x 350
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......