Proline Kryomats
YAUE0007 / 10/2/2018
Starting up
43
The momentary bath temperature is displayed,
the pump starts provided "Standby" or "Manual start" (
7.6.1) has
not been programmed,
all values are accepted which were active before switch-off.
Check or set overtemperature cut-off point:
The switching point is shown in the LED display on pressing the key
.
Change overtemperature cut-off (
7.12.1). Overtemperature
protection and checking on page 89.
If necessary, top up heat transfer liquid, this has been pumped out
by filling the external load.
Display for
LEUEL
(low level) appears when the bath has too
little liquid.
Red LED
above the fault triangle
flashes.
Find cause of fault and, where necessary, top up missing heat
transfer liquid (
Press the Enter key.
Also, press the key if unit has been switched off in the fault state.
No release is possible on Command remote control!
Command
Language
If the Command remote control is being
switched on for the first time, the
illustrated window appears automatically,
enabling you to select the dialog language
with the appropriate soft key.
Bath temperature
02%32
°C
Level alarm
LEUEL
°C
Overtemp. cut-off
90
°C
English
Deutsch
Français
Español
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......