Proline Kryomats
118
Maintenance
/ 10/2/2018
9.3.6
Remedying faults
Before you contact the LAUDA Service Temperature Equipment, check whether the problem can be
remedied with the following instructions:
Fault
Possible remedy
Device does not cool or only very slowly.
1. The module "Smart Cool" is set to "off"
Switch
on "Smart Cool" module
2. Dirty condenser
Clean condenser
3. Temperature limit Til too high
Reduce
temperature limit Til
Device does not heat up or only very slowly.
Temperature limit Tih too low
Increase temperature
limit Tih
The compressors are running although there
is no necessity of cooling.
Regular device function (protectiv function).
Master: Alarm message
tEMNP
Command: Overtemperature protection.
Wait until the outflow temperature has cooled below
the overtemperature cut-off point or set the cut-off
point higher than the outflow temperature.
Master: Warning message
WXArn
104
Command: Level very low
(Imminent low level in the bath vessel).
Master: Alarm message
LEUEL
Command: Low level.
(Low level in the bath vessel)
1. Check hoses, connections and load for whether a
leaky location is present.
As applicable, rectify
the leakage and top up the missing heat carrier
liquid
2. Check the Proline Kryomat for whether a leaky
location is present.
If necessary, contact
LAUDA Service Constant Temperature
Equipment
3. The liquid may drop due to cooling or degassing.
If necessary, top up the missing heat transfer
liquid
Master: Warning message
WXArn
103
Command: Level too high
(Imminent excessive level in the bath
vessel).
Master: Alarm message
AL 6
Command: Level too high
(Excessive level in the expansion vessel)
1. Volume expansion of the heat transfer liquid
during heating up.
2. Moisture absorption in the heat transfer liquid.
Master: Alarm message
bl0C
Command: Pump blocked
(Pump motor monitoring: Overload,
blockage).
1. The viscosity of the heat teransfer liquid is too
high
change heat carrier liquid or raise the
setpoint temperature.
2. The pump is blocked.
Contact the LAUDA
Service Constant Temperature Equipment
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...