Proline heating thermostat
41 / 123
Command
T
ext
Provided an external temperature probe is con-
nected, its value is displayed in the lower left part
of the standard and super windows (applies to the
works setting for the window partitioning).
External actual temperatures can also be read in
via interface modules. (
8).
7.7.3
Setting pump power or standby
With the Proline Varioflex pump, 8 pump levels are available with which the bath circulation, flow rate and pressure, the
noise generated and the mechanical heat input can be optimized. This is particularly advantageous with coolers. With
smaller coolers (e.g. P8) without an external load, Power Level 3 to 4 is practicable and sufficient.
Master
Pu
and 1 x
Call pump power levels display
Pu
.
The current pump level is displayed (here
5
).
The pump levels display flashes.
or
Select pump level (pump speed = pump power):
1
to
8
for pump operation. Pump responds immediately!
0
activates the standby function (pump and heater are deactivated).
Wait 4 seconds or
Display flashes 4 s
new value is automatically accepted, or
value is immediately accepted with Enter key.
Pu 5
°C
Pump Menu Screen
T
set
T
fix
T
int
°C
25,
01
T
set
°C
25,
00
Level
4
T
ext
°C
25,
02
Pump
3
Summary of Contents for PROLINE P 5 C
Page 13: ...02 2022 Proline heating thermostat 13 123 Clear view thermostat PVL 24 Bridge thermostat PB C ...
Page 113: ...02 2022 Proline heating thermostat 113 123 Pump characteristics measured with water ...
Page 114: ...114 123 Proline heating thermostat 02 2022 Pump characteristics measured with water ...
Page 115: ...02 2022 Proline heating thermostat 115 123 12 Declaration of Conformity ...
Page 118: ......
Page 119: ......