7.13.2 External control variable (External measurement probe)
You only need to read further here if you have connected an external temperature probe or the actual
temperature is read in from a module (and you have activated it as control variable according to (
Sec-
tion 7.7.4)).
Only modify the control parameters if you have knowledge of control techniques.
The control system for external actual values is implemented for improvement of the control behavior as
a two-stage cascade controller. A "master controller" determines the "internal setpoint", from the tem-
perature setpoint and the external temperature, passed to the slave controller. The control value of the
slave controller controls the heating.
When a setpoint step change is specified, it may be that the optimum control would set a bath tempera-
ture, which might significantly exceed the temperature desired on the external vessel. There is a correc-
tion limitation, which specifies the maximum permissible deviation between the temperature on the ex-
ternal load and the bath temperature.
Master
P1d..
3x
to
MNEnu
1x
to
P1d..
As shown in the menu structure (
7.6), the parameters for the ex-
ternal control variable can now be set.
Select the parameter with
or
and confirm with
. The set
value is displayed. Adapt it with
or
confirm with
.
Example: Proportional factor of the master controller Kpe = 1.5.
Parameters
Ma
ster controller (PIDT
1
-controller):
Ma proportional factor:
EP
= Kpe
as factor.
Ma proportional range:
Eb
= Prop_E in Kelvin.
Ma reset time:
En
= Tne
in seconds.
Ma derivative time (auto/ man)
Ev
= Tve
in seconds.
The thermostat logic system only permits values with Tne > Tve!
Ma damping time (auto/ man)
Ed
= Tde
in seconds.
Parameter
Sl
ave controller (P-controller):
Sl proportional range:
iP
= Xpf
in Kelvin.
Tve, Tde, Prop_E changeover to Auto/ Man
E
A
= Automat-
ic or
MN
= Manual.
Works setting is Automatic. These three parameters should only be
modified by experienced control technicians!
Hold
pressed for 3s, then make changes.
Correction limitation
Et0l
. See introduction (
EP !5
°C
Et0l
°C
E A
°C